Supply of Laboratory and/or Medical Equipment (Gel Documentation System & Electrostatic Field Meter & Multiparameter Analyzer & Dynamic Electrosleep Apparatus) for SES Uzbekistan
UNOPS
Supply of Laboratory and/or Medical Equipment (Gel Documentation System & Electrostatic Field Meter & Multiparameter Analyzer & Dynamic Electrosleep Apparatus) for SES Uzbekistan
Request for quotation
Reference:
RFQ/2026/63454
Beneficiary countries or territories:
Uzbekistan
Registration level:
Basic
Published on:
22-Jun-2026
Deadline on:
14-Jul-2026 07:00 0.00
Expired
Description
This RFQ refers to the Supply of Laboratory Equipment for SES of the Republic of Uzbekistan, divided into 4 lots:
Lot #1: Gel Documentation System - 1 unit
Lot #2:Electrostatic field meter, portable - 2 units
Lot #3: Multiparameter analyzer (anemometer, luxmeter, lightmeter, UVmeter) - 3 units
Lot #4: Dynamic electrosleep apparatus - 1 unit
Note to Bidders: Bidders are allowed to quote prices for one or more lots identified in this tender. However, the Bidders must offer 100% of the quantities specified for each lot. Evaluation will be done separately for each lot.
IMPORTANT NOTE: Interested vendors must respond to this tender using the UNOPS eSourcing system, via the UNGM portal. In order to access the full UNOPS tender details, request clarifications on the tender, and submit a vendor response to a tender using the system, vendors need to be registered as a UNOPS vendor at the UNGM portal and be logged into UNGM. For guidance on how to register on UNGM and submit responses to UNOPS tenders in the UNOPS eSourcing system, please refer to the user guide and other resources available at: https://esourcing.unops.org/#/Help/Guides
Interested in improving your knowledge of what UNOPS procures, how we procure and how to become a vendor to supply to our organization? Learn more about our free online course on “Doing business with UNOPS” here
Lot #1: Gel Documentation System - 1 unit
Lot #2:Electrostatic field meter, portable - 2 units
Lot #3: Multiparameter analyzer (anemometer, luxmeter, lightmeter, UVmeter) - 3 units
Lot #4: Dynamic electrosleep apparatus - 1 unit
Note to Bidders: Bidders are allowed to quote prices for one or more lots identified in this tender. However, the Bidders must offer 100% of the quantities specified for each lot. Evaluation will be done separately for each lot.
IMPORTANT NOTE: Interested vendors must respond to this tender using the UNOPS eSourcing system, via the UNGM portal. In order to access the full UNOPS tender details, request clarifications on the tender, and submit a vendor response to a tender using the system, vendors need to be registered as a UNOPS vendor at the UNGM portal and be logged into UNGM. For guidance on how to register on UNGM and submit responses to UNOPS tenders in the UNOPS eSourcing system, please refer to the user guide and other resources available at: https://esourcing.unops.org/#/Help/Guides
Interested in improving your knowledge of what UNOPS procures, how we procure and how to become a vendor to supply to our organization? Learn more about our free online course on “Doing business with UNOPS” here
This tender has been posted through the UNOPS eSourcing system. / Cet avis a été publié au moyen du système eSourcing de l'UNOPS. / Esta licitación ha sido publicada usando el sistema eSourcing de UNOPS. Vendor Guide / Guide pour Fournisseurs / Guíra para Proveedores: https://esourcing.unops.org/#/Help/Guides
First name:
N/A
Surname:
N/A
This procurement opportunity integrates considerations for at least one sustainability indicator. However, it does not meet the requirements to be considered sustainable.
Sustainable resource use
Environmental
The tender contains sustainability considerations promoting the sustainable use of resources.
Examples:
Energy-saving measures, recycling, take-back programmes and responsible end-of-life management
| Link | Description | |
|---|---|---|
| https://esourcing.unops.org/#/Help/Guides | UNOPS eSourcing – Vendor guide and other system resources / Guide pour fournisseurs et autres ressources sur le système / Guía para proveedores y otros recursos sobre el sistema |
New clarification added: Q8: Reference is made to Lot 3: Multiparameter analyzer technical requirement line C:Power supply. Please clarify if a) 220 VAC mains power will be available at the installation location; or b) the system must operate using a solar panel solution. A8:The instrument shall include a rechargeable battery for autonomous operation. A charger and/or AC power adapter compatible with 220–240 VAC, 50 Hz mains supply with a European standard plug shall be provided. Solar power supply is not required.
Edited on:
10-Jul-2026 14:26
Edited by:
webservice@unops.org
New clarification added: Q7: Reference is made to Lot 3: Multiparameter analyzer technical requirement lines #10 and #11 on data logger. Please clarify the following: Will operators simply visit the installation site and manually read the values on the display? Is a remote access via Wi-Fi, LAN, or a cellular modem required? Should the system automatically send data to an FTP server or via MQTT protocol for subsequent analysis using dedicated software? Alternatively, is there intention to connect the system to an existing LAN or Wi-Fi network and access the data through software installed on a PC connected to the same network? A7:The equipment is intended as a portable measuring instrument operated by personnel at the measurement location. Therefore remote access via Wi-Fi, LAN, cellular modem, FTP, MQTT or similar technologies is not required. Please follow the minimum required functionalities specified. However you can offer a device with wider functionalities.
Edited on:
10-Jul-2026 14:21
Edited by:
webservice@unops.org
New clarification added: Q6: Reference is made to Lot 3: Multiparameter analyzer technical requirement line # 8: Air velocity measurement range: at least 0.1 m/s to 20 m/s. Would models that are capable of measuring significantly higher wind speeds and that include an integrated weather sensor capable of measuring both Wind speed and Wind direction be acceptable. For this point, clarify whether there is a requirement to provide a cup anemometer, typically used for outdoor meteorological measurements or a hot-wire anemometer, typically used for indoor airflow monitoring and ventilation studies. A6: The intended application is the measurement of air movement for environmental and occupational hygiene purposes. Accordingly, the specified measurement range is sufficient. A hot-wire or another equivalent/better technology capable of meeting the specified measurement range and accuracy is acceptable.
Edited on:
10-Jul-2026 14:16
Edited by:
webservice@unops.org
New clarification added: Q5: Reference is made to Lot 3: Multiparameter analyzer technical requirement lines # 6 and #7 on temperature and humidity measurements respectively. If installation of the equipment is outdoors, it is recommend to include a radiation shield (radiant screen) to ensure accurate measurements and protect the sensors from direct solar exposure. A5:The equipment will be used as a portable measuring instrument. Therefore, a radiation shield is not required.
Edited on:
10-Jul-2026 14:11
Edited by:
webservice@unops.org
New clarification added: Q4: Reference is made to Lot 3: Multiparameter analyzer technical requirement line #5: UV irradiance spectral measurement range: 280–400 nm region with an accuracy: ±10 % of reading or better. This implies the need to measure both UV-A and UV-B to cover this range accurately, two dedicated sensors are usually offered, one for each spectral band. UV-A and UV-B measurements are also generally associated with outdoor environmental monitoring, which reinforces the need to clarify the installation environment. A4:Please note that required device is intended for portable field and laboratory measurements rather than continuous outdoor environmental monitoring. The offered solution may consist of either a single sensor covering the required spectral range or two dedicated sensors (UV-A and UV-B), provided together they cover the required measurement range.
Edited on:
10-Jul-2026 14:07
Edited by:
webservice@unops.org
New clarification added: Q3: Reference is made to Lot 3: Multiparameter analyzer technical requirement line #3: Illuminance measurement range: at least 10 lx to 200,000 lx with resolution 1lx or better and accuracy ±8% of reading or better. 1 lux resolution over the entire range from 10 lux to 200,000 lux is highly unlikely from a technical perspective, and this requirement may need to be reviewed considering practical sensor capabilities. Would it be considered to amend the requirement as follows: Measurement range: 0 to 150,000 lux, Resolution: 10 lux, Accuracy: ±3%. A3:The specified illuminance measurement range and resolution will remain unchanged. However, please note that if an offered solution provides minor deviations from our requirements while remaining suitable for the intended application, such deviations may be considered technically acceptable during the technical evaluation, subject to confirmation by the End User.
Edited on:
10-Jul-2026 14:03
Edited by:
webservice@unops.org
New amendment added #2: The RFQ has been further amended in order to:1) replace document titled "RFQ_2026_63454_Section III_Returnable Bidding Forms_Form C_Technical Bid Form" with "RFQ_2026_63454_Section III_Returnable Bidding Forms_Form C_Technical Bid Form_REV1_06JUL26" that introduces amendments (in red font) made to several technical requirement lines;2) extend quotation submission deadline from 09-Jul-2026 till 14-Jul-2026
Edited on:
06-Jul-2026 10:13
Edited by:
webservice@unops.org
New clarification added: Q2: The reference is made to Lot #1: Gel Documentation System and further to the technical requirement lines #9. Maximum gel size, cm: at least 28 × 36 and #10. Maximum scanning area size, cm: at least 19.4 × 26. Based on a review of currently available gel documentation systems from leading manufacturers, the requirement for Maximum gel size reflects the specifications of older-generation systems and is no longer a standard parameter used by most manufacturers of modern gel documentation systems. Current-generation systems are typically specified by their maximum imaging (scanning) area, which defines the effective area available for image acquisition and analysis. In addition, modern imaging systems generally feature more compact imaging areas while providing significantly improved sensitivity, image quality, dynamic range, and software capabilities. The currently specified maximum scanning area therefore significantly limits the range of current-generation systems available on the market without providing a corresponding analytical advantage. Therefore it is requested a) to remove technical requirment line #9 cal specification and b) to revise technical requirement #10 as "... at least 20 × 16 cm, which is sufficient for imaging standard agarose and polyacrylamide gels used in routine molecular biology laboratories while allowing participation of current-generation gel documentation systems that provide equivalent or superior analytical performance. A2: The request to remove technical requirement #9. Maximum gel size, cm: at least 28 × 36 is accepted. As for technical requirement #10. Maximum scanning area size, cm: at least 19.4 × 26, it will be revised to at least 18 × 20 cm, as per the end-user, to allow wider competition. The RFQ_2026_63454_Section III_Returnable Bidding Forms_Form C_Technical Bid Form will be updated accordingly and uploaded in the "Documents" section of the RFQ.
Edited on:
06-Jul-2026 08:52
Edited by:
webservice@unops.org
New amendment added #1: This RFQ is hereby amended as follows:The deadline for submission of bids is extended from 02 July 2026, 07:00 UTC to 09 July 2026, 07:00 UTC.The deadline for submission of clarification requests is extended from 26 June 2026, 07:00 UTC to 03 July 2026, 07:00 UTC.
Edited on:
25-Jun-2026 08:28
Edited by:
webservice@unops.org
New clarification added: Q1: Extension of the original deadline from July 02, 2026 to July 09, 2026 is requestedA1: The request on extension will be accommodated accordinly.
Edited on:
25-Jun-2026 08:17
Edited by:
webservice@unops.org