Supply of Laboratory Equipment (Part CXVI) for SES of the Republic of Uzbekistan.
UNOPS
Supply of Laboratory Equipment (Part CXVI) for SES of the Republic of Uzbekistan.
Request for quotation
Reference:
RFQ/2026/62474
Beneficiary countries or territories:
Uzbekistan
Registration level:
Basic
Published on:
27-Apr-2026
Deadline on:
12-May-2026 07:00 0.00
Description
Tender description: This RFQ refers to the Supply of Laboratory Equipment (Part CXVI) for SES of the Republic of Uzbekistan. The requirement consist of 2 items in the following 2 Lots:
Lot 1# Digital X-Ray, fixed- 1 Units
Lot 2# Mobile X- Ray- 2 Units
Note to Bidders: Bidders are allowed to quote prices for one or more lots identified in this tender. However, the Bidders must offer 100% of the quantities specified for each lot. Evaluation will be done separately for each lot.
IMPORTANT NOTE: Interested vendors must respond to this tender using the UNOPS eSourcing system, via the UNGM portal. In order to access the full UNOPS tender details, request clarifications on the tender, and submit a vendor response to a tender using the system, vendors need to be registered as a UNOPS vendor at the UNGM portal and be logged into UNGM. For guidance on how to register on UNGM and submit responses to UNOPS tenders in the UNOPS eSourcing system, please refer to the user guide and other resources available at: https://esourcing.unops.org/#/Help/Guides
Interested in improving your knowledge of what UNOPS procures, how we procure and how to become a vendor to supply to our organization? Learn more about our free online course on “Doing business with UNOPS” here
Lot 1# Digital X-Ray, fixed- 1 Units
Lot 2# Mobile X- Ray- 2 Units
Note to Bidders: Bidders are allowed to quote prices for one or more lots identified in this tender. However, the Bidders must offer 100% of the quantities specified for each lot. Evaluation will be done separately for each lot.
IMPORTANT NOTE: Interested vendors must respond to this tender using the UNOPS eSourcing system, via the UNGM portal. In order to access the full UNOPS tender details, request clarifications on the tender, and submit a vendor response to a tender using the system, vendors need to be registered as a UNOPS vendor at the UNGM portal and be logged into UNGM. For guidance on how to register on UNGM and submit responses to UNOPS tenders in the UNOPS eSourcing system, please refer to the user guide and other resources available at: https://esourcing.unops.org/#/Help/Guides
Interested in improving your knowledge of what UNOPS procures, how we procure and how to become a vendor to supply to our organization? Learn more about our free online course on “Doing business with UNOPS” here
This tender has been posted through the UNOPS eSourcing system. / Cet avis a été publié au moyen du système eSourcing de l'UNOPS. / Esta licitación ha sido publicada usando el sistema eSourcing de UNOPS. Vendor Guide / Guide pour Fournisseurs / Guíra para Proveedores: https://esourcing.unops.org/#/Help/Guides
First name:
N/A
Surname:
N/A
This procurement opportunity integrates considerations for at least one sustainability indicator. However, it does not meet the requirements to be considered sustainable.
Sustainable resource use
Environmental
The tender contains sustainability considerations promoting the sustainable use of resources.
Examples:
Energy-saving measures, recycling, take-back programmes and responsible end-of-life management
| Link | Description | |
|---|---|---|
| https://esourcing.unops.org/#/Help/Guides | UNOPS eSourcing – Vendor guide and other system resources / Guide pour fournisseurs et autres ressources sur le système / Guía para proveedores y otros recursos sobre el sistema |
42201806
-
Medical radiology and fluoroscopy RF units
New clarification added: Q5: LOT 1 – Mobile X-ray System, Please clarify whether the requirements under Specifications B.5, B.9, B.10, B.20, B.24, B.25, B.28, and B.37 may be revised as proposed?: Q5(i): Specification B.5 (Column height for transportation), please clarify whether the maximum column height requirement may be revised from 1250 mm to 1400 mm.A5(i): This parameter remains unchanged. However, certain deviations may be considered technically acceptable, subject to confirmation by the end user. Q5(ii): Specification B.9 (Arm extension stroke range), please clarify whether the minimum required range may be revised from 550 mm to 350 mm. A5(ii): A reduction of the arm extension stroke to 350 mm is not considered acceptable, as it significantly limits positioning flexibility required for a broad range of radiographic examinations. Q5(iii): Specification B.10 (Maximum arm length when extended), please clarify whether the minimum required length may be revised from 1200 mm to 1000 mm. A5(iii): This parameter remains unchanged. However, certain deviations may be considered technically acceptable, subject to confirmation by the end user. Q5(iv): Specification B.20 (Driving speed), please clarify whether the requirement may be revised to specify forward driving speed not less than 5.0 km/h. A5(iv): This parameter remains unchanged. However, certain deviations may be considered technically acceptable, subject to confirmation by the end user. Q5(v): Specification B.24 (Main body design), please clarify whether the requirement for a flat and wide top surface for placement of a laptop PC may be removed or revised, allowing systems with integrated touchscreen consoles and fully digital workflow (DICOM compliant, RIS/PACS connectivity). A5(v): This parameter remains unchanged. However, certain deviations may be considered technically acceptable, subject to confirmation by the end user. Q5(vi): Specification B.25 (FPD anti-theft lock), please clarify whether the requirement may be revised to allow a system-level locking mechanism instead of a detector-specific anti-theft lock, or otherwise be adjusted accordingly. A5(vi): This parameter remains unchanged. However, certain deviations may be considered technically acceptable, subject to confirmation by the end user. Q5(vii): Specification B.28 (kV setting range), please clarify whether the required range may be revised from 40–133 kV to 40–125 kV, provided that image quality and clinical performance requirements are maintained. A5(vii): This parameter remains unchanged. However, certain deviations may be considered technically acceptable, subject to confirmation by the end user. Q5(viii): Specification B.37 (Collimation controls), please clarify whether the requirement for controls on both front and rear sides may be revised to allow controls on the front side only, provided that safe and ergonomic operation is ensured. A5(viii): This parameter remains unchanged. However, certain deviations may be considered technically acceptable, subject to confirmation by the end user.
Edited on:
07-May-2026 11:09
Edited by:
webservice@unops.org
New clarification added: Q4: LOT 1 – Digital X-ray System, S/N I-54 (Tomosynthesis / Tomography): Please clarify whether systems with integrated tomography capabilities may be accepted as an alternative to tomosynthesis, provided that equivalent or superior clinical performance is achieved, and whether related additional hardware requirements may be adjusted accordingly. A4: Modern DR/DR-RF systems widely utilize tomosynthesis and other advanced reconstruction techniques, which provide clinically sufficient sectional imaging and effective reduction of tissue overlap. Therefore, this parameter remains unchanged.
Edited on:
07-May-2026 11:02
Edited by:
webservice@unops.org
New clarification added: Q3: LOT 1: Digital X-ray System, S/N G-37 (Raster Ratio), please clarify whether a raster ratio of not less than 10:1 may be accepted. A3: The specified anti-scatter grid ratio is a widely adopted standard in high-performance radiographic and fluoroscopic systems, ensuring optimal scatter rejection and improved image contrast. Therefore, this parameter remains unchanged. However, bidders offering a lower raster ratio are still encouraged to submit bids. Deviations from the technical requirements may be considered technically acceptable, subject to confirmation by the end user.
Edited on:
07-May-2026 11:01
Edited by:
webservice@unops.org
New clarification added: Q2: LOT 1: Digital X-ray System, S/N F-32 (Acquisition Formats), please clarify whether the requirement may be revised to a minimum of three (3) acquisition formats, other than the full format. A2: Modern flat panel detector systems provide multiple acquisition formats to ensure clinical workflow flexibility. The specified requirement reflects a standard configuration in commercially available high-end DR systems and therefore remains unchanged. However, bidders offering fewer acquisition formats are still encouraged to submit bids. Deviations from the technical requirements may be considered technically acceptable, subject to confirmation by the end user.
Edited on:
07-May-2026 11:01
Edited by:
webservice@unops.org
New clarification added: Q1: LOT 1: Digital X-ray System, S/N E-28 (Filtration), Please clarify whether manufacturer-specific filtration configurations (e.g., combinations of Cu and/or Al such as 0.1 mm Cu, 0.2 mm Cu, 0.3 mm Cu, 1–2 mm Al, or equivalent) may be accepted, provided they comply with applicable international standards and ensure the required clinical performance. A1: The specified filtration configuration reflects current international standards for digital radiography and fluoroscopy systems and is supported by multiple leading manufacturers. Therefore, this parameter remains unchanged. However, if the proposed equipment uses an alternative filtration configuration equivalent to the specified requirement, bidders are still encouraged to submit their bids. Proposals containing deviations from the technical requirements may be considered technically acceptable, subject to confirmation by the end user.