Supply, delivery and installation of Laboratory Equipment to Health Facilities in Uzbekistan
UNOPS
Supply, delivery and installation of Laboratory Equipment to Health Facilities in Uzbekistan
Request for quotation
Reference:
RFQ/2023/45672
Beneficiary countries or territories:
Uzbekistan
Registration level:
Basic
Published on:
29-May-2023
Deadline on:
03-Jul-2023 10:00 0.00
Description
This RFQ refers to the supply, delivery and installation of Laboratory Equipment to various hospitals across Uzbekistan. The requirement consists of the following 34 lots:
Bidders are allowed to quote prices for one or more lots identified in this tender. However, the Bidders must offer 100% of the quantities specified for each lot. Evaluation will be done separately for each lot. UNOPS reserves the right to vary the quantities of lots or remove a lot, and award multiple contracts for each lot.
IMPORTANT NOTE: Interested vendors must respond to this tender using the UNOPS eSourcing system, via the UNGM portal. In order to access the full UNOPS tender details, request clarifications on the tender, and submit a vendor response to a tender using the system, vendors need to be registered as a UNOPS vendor at the UNGM portal and be logged into UNGM. For guidance on how to register on UNGM and submit responses to UNOPS tenders in the UNOPS eSourcing system, please refer to the user guide and other resources available at: https://esourcing.unops.org/#/Help/Guides
- Blood Gas & Electrolytes Analyzer -17 units;
- Automated Hematology Analyzer - 17 units;
- Automated Biochemical Analyzer - 17 units;
- Automated Immunoassay Analyzer - 1 unit;
- Automated Microbial Identification Analyzer - 1 unit;
- Flow Cytometer and Reagents – 4 units;
- Double Beam Spectrophotometer – 1 unit;
- Single-Beam Luminescent Spectrometer - 1 unit;
- Atomic Absorption Spectrophotometer - 1 unit;
- Liquid Chromatograph - 1 unit;
- Automated Polarimeter - 1 unit;
- Fraction Collector - 1 unit;
- PCR analyzer, real-time - 1 unit;
- Gene amplification detector (OSNA method) - 1 unit;
- Microarray scanner - 1 unit;
- Laminar cabinet - 2 units;
- Cytogenetic and Molecular equipment lab - 1 unit;
- Anatomic pathology with IHC (immunochemistry) lab consumables - 1 unit
- Cryostat and reagent (frozen section) - 4 units;
- Plasma collection system - 1 unit
- Therapeutic membrane plasmapheresis system - 2 units;
- Automated tissue processor - 16 units;
- Paraffin embedding station - 16 uits;
- Slide dryer - 16 units;
- Semi-automated microtome - 16 units;
- Biological samples cryogenic storage system - 2 units;
- Rotary evaporator - 2 units;
- Lab Thermostat - 17 units;
- Refrigerated Centrifuge - 3 units;
- Automatic Blood Cell Separator - 1 unit;
- CO2 Incubator - 1 unit;
- Ultra-low Freezer (not less than -86°С) - 2 units;
- Analytical Lab Balance - 2 units;
- Lab Vortex Mixer - 5 units.
Bidders are allowed to quote prices for one or more lots identified in this tender. However, the Bidders must offer 100% of the quantities specified for each lot. Evaluation will be done separately for each lot. UNOPS reserves the right to vary the quantities of lots or remove a lot, and award multiple contracts for each lot.
IMPORTANT NOTE: Interested vendors must respond to this tender using the UNOPS eSourcing system, via the UNGM portal. In order to access the full UNOPS tender details, request clarifications on the tender, and submit a vendor response to a tender using the system, vendors need to be registered as a UNOPS vendor at the UNGM portal and be logged into UNGM. For guidance on how to register on UNGM and submit responses to UNOPS tenders in the UNOPS eSourcing system, please refer to the user guide and other resources available at: https://esourcing.unops.org/#/Help/Guides
This tender has been posted through the UNOPS eSourcing system. / Cet avis a été publié au moyen du système eSourcing de l'UNOPS. / Esta licitación ha sido publicada usando el sistema eSourcing de UNOPS. Vendor Guide / Guide pour Fournisseurs / Guíra para Proveedores: https://esourcing.unops.org/#/Help/Guides
First name:
N/A
Surname:
N/A
This procurement opportunity integrates considerations for at least one sustainability indicator. However, it does not meet the requirements to be considered sustainable.
Prevention of pollution
Environmental
The tender contains sustainability considerations for the prevention of polluting emissions to air, solid waste to land and discharges to water.
Examples:
EMS, waste management and wastewater management.
| Link | Description | |
|---|---|---|
| https://esourcing.unops.org/#/Help/Guides | UNOPS eSourcing – Vendor guide and other system resources / Guide pour fournisseurs et autres ressources sur le système / Guía para proveedores y otros recursos sobre el sistema |
95141902
-
Laboratory unit
41101502
-
Stomachers
41101503
-
Laboratory sprayers
41101504
-
Homogenizers
41101515
-
Liquid measuring cans
41101516
-
Dounce homogenizers
41101518
-
Laboratory blenders or emulsifiers
41101519
-
Laboratory cell disruptor
41101701
-
Laboratory mills
41101703
-
Tissue grinders
41101705
-
Laboratory crushers or pulverizers
41101706
-
Laboratory disintegrators
41101707
-
Laboratory presses
41101708
-
Laboratory grinder or polisher
41101709
-
Laboratory asphalt and concrete mixer
41101801
-
Electron guns
41101802
-
X ray generators
41101803
-
Coulometers
41101804
-
Electroscopes
41101805
-
Fluxmeters
41101806
-
Magnetometers
41101807
-
Electron diffraction apparatus
41101808
-
Neutron diffraction apparatus
41101809
-
Optical diffraction apparatus
41101810
-
Diffractometers
41101811
-
Electron probe x ray micro analyzer
41101812
-
Particle accelerator
41101813
-
Laboratory X ray equipment controller
41101901
-
Ion sources
41101902
-
Ion exchange apparatus
41101903
-
Ion implantation equipment
41102401
-
Gas burners
41102402
-
Spirit burners
41102403
-
Laboratory incinerators
41102404
-
Laboratory heaters
41102405
-
Heating mantles or tapes
41102406
-
Laboratory hotplates
41102407
-
Warming cabinets
41102410
-
Infrared dryers
41102412
-
Hot air blowers
41102421
-
Temperature cycling chambers or thermal cyclers
41102422
-
Dry baths or heating blocks
41102423
-
Stirring hotplates
41102424
-
Slide warmers
41102425
-
Slide dryers
41102426
-
Heating or drying equipment or accessories
41102427
-
Drying tower
41102428
-
Bibulous paper
41102429
-
Blood unit tubing heat sealer
41102430
-
Laboratory general purpose heat sealer
41102501
-
Laboratory insect containers
41102502
-
Rearing facilities for entomology
41102503
-
Fabric or netting for entomology
41102504
-
Entomological pinning equipment
41102505
-
Entomological mounting materials
41102506
-
Entomological trays
41102507
-
Entomological catching equipment
41102508
-
Entomological aspirators
41102509
-
Entomological dippers
41102510
-
Entomological monocups
41102511
-
Entomological sticky traps
41102512
-
Insect test kits
41102513
-
Entomological display units
41102601
-
Laboratory cages for small animals
41102602
-
Aquaria equipment
41102603
-
Animal identification supplies
41102604
-
Animal catching devices
41102605
-
Fish aeration systems
41102606
-
Laboratory animal restraints or harnesses
41102608
-
Animal testing equipment
41102610
-
Animal for research testing
41102611
-
Research animal food and diet
41102612
-
Research animal induction chamber
41102613
-
Research animal physiological test kit
41102614
-
Research animal bedding material
41102701
-
Crystal lattice models
41102702
-
Scintillation crystal assemblies
41102703
-
Light scattering equipment
41102704
-
X ray diffraction equipment
41102705
-
Crystallizers
41102901
-
Tissue embedding stations
41102905
-
Histological staining apparatus
41102909
-
Tissue processors
41102910
-
Tissue culture apparatus
41102911
-
Histological knives or knife holders or blades
41102912
-
Histological glass knife makers
41102913
-
Histological hones or straps or compounds
41102914
-
Ultrasonic disintegrators
41102915
-
Histology sampling and dissecting stations
41102916
-
Microtomes
41102917
-
Microtome blades
41102918
-
Laboratory cover slippers
41102919
-
Solvent recyclers
41102920
-
Histology tissue cassettes
41102921
-
Histology paraffin
41102922
-
Automated cover slipping equipment
41102923
-
Somatic cell counter
41102928
-
Automated microscope slide labeler
41103001
-
Refrigerated cooling plate probes
41103003
-
Cryostats
41103004
-
Fan circulated ovens
41103005
-
Ultra cold or ultralow upright cabinets or freezers
41103006
-
Cryogenic or liquid nitrogen freezers
41103007
-
Chilling units or cold water circulators
41103008
-
Refrigerated cooling modules
41103010
-
Blood bank refrigerators
41103011
-
General purpose refrigerators or refrigerator freezers
41103012
-
Flammable material storage refrigerators or refrigerator freezers
41103013
-
Explosion proof refrigerators or refrigerator freezers
41103014
-
Chromatography refrigerators
41103015
-
Blood bank freezers
41103017
-
Flammable material storage freezers
41103019
-
Plasma storage freezers
41103020
-
Ultra cold or ultralow chest freezers
41103021
-
Laboratory plate freezers
41103022
-
Cool transport or storage
41103023
-
Laboratory chillers
41103024
-
Cold traps
41103025
-
Laboratory cooling equipment accessories
41103026
-
Benchtop ice bucket or chilling container
41103027
-
Blood unit storage boot
41103028
-
Liquid nitrogen measuring stick
41103029
-
Cryogenic storage cane
41103030
-
Cryogenic storage bag or overwrap
41103031
-
Insulated transport cooler or tote
41103032
-
Cryogenic tube or vial permanent marker or label
41103033
-
Laboratory dewar flask
41103034
-
Cryobead system for microbial organism cryogenic storage
41103035
-
Fresh frozen plasma storage carton or frame
41103036
-
Refrigerated specimen storage rack or tray
41103037
-
Polymerase chain reaction PCR tube strip and plate cooler
41103201
-
Chemical engineering washers
41103202
-
Laboratory washing machines
41103203
-
Pipette washers
41103205
-
Washing machine racks or accessories
41103206
-
Laboratory washing detergents
41103207
-
Microplate washers
41103208
-
Blood bank cell washers
41103209
-
Laboratory wash bottles
41103210
-
Laboratory ultraviolet UV sterilizers
41103211
-
Laboratory wastewater treatment equipment
41103212
-
Lab glassware cleaning brush
41103213
-
Laboratory surface and glassware decontaminant
41103301
-
Liquid scintillation counters
41103302
-
Battery acid hydrometers
41103303
-
Densitometers
41103305
-
High vacuum equipment
41103306
-
Pneumatic vacuum equipment
41103307
-
Vacuum or mercury vapour equipment
41103308
-
High vacuum combustion apparatus
41103309
-
Flow injection analysis equipment
41103310
-
Gas or vapour concentration measuring instruments
41103311
-
Manometers
41103312
-
Viscosimeters
41103313
-
Depth indicators
41103314
-
Microscopic structure estimation apparatus
41103315
-
Solution strength estimation apparatus
41103316
-
Pycnometers
41103317
-
Surface tension measuring instruments
41103318
-
Nuclear densitometer
41103319
-
Concentration measurement instrument
41103320
-
Density measurement instrument
41103321
-
Floating body stability measurement apparatus
41103322
-
Draft gauge
41103323
-
Open channel acoustic flowmeter
41103324
-
Laboratory wave generator
41103325
-
Visual fluid flow apparatus
41103326
-
Wind tunnel
41103327
-
Current meter
41103328
-
Vacuum based pipette aspirator system
41103401
-
Contamination control screens
41103403
-
Microbiological aircontrol equipment
41103406
-
Isolation glove boxes
41103407
-
Anaerobic chamber
41103408
-
Refrigerated reach in environmental or growth chambers
41103409
-
Heated reach in environmental or growth chambers
41103410
-
Refrigerated and heated reach in environmental or growth chambers
41103411
-
Refrigerated walk in environmental or growth chambers
41103412
-
Heated walk in environmental or growth chambers
41103413
-
Refrigerated and heated walk in environmental or growth chambers
41103414
-
Laboratory environmental conditioning equipment accessories
41103415
-
Clean benches
41103416
-
Temperature cycle chamber
41103417
-
Indoor air quality monitor
41103418
-
Temperature and humidity walk in environmental chamber
41103419
-
Pollution environmental chamber
41103420
-
Ozone environmental chamber
41103421
-
Explosion environmental chamber
41103501
-
Ebuliometer
41103502
-
Fume hoods or cupboards
41103504
-
Laminar flow cabinets or stations
41103506
-
PCR enclosures
41103507
-
HEPA filtered enclosures
41103508
-
Carbon filtered enclosures
41103509
-
Laboratory scrubbers
41103510
-
Laboratory blowers
41103511
-
Laboratory enclosure accessories
41103512
-
Static eliminators
41103513
-
Tissue culture enclosures
41103514
-
Laboratory steam generator
41103515
-
Laboratory gas generator
41103516
-
Arm hood
41103517
-
Biological safety cabinet
41103701
-
Circulating baths
41103702
-
Thermostatic baths
41103703
-
Multiple baths
41103704
-
Biological baths
41103705
-
Organ baths
41103706
-
Water baths
41103707
-
Oil baths
41103708
-
Sand baths
41103709
-
Refrigerated baths
41103710
-
Orbital shaking water baths
41103711
-
Reciprocating shaking water baths
41103712
-
Immersion circulators
41103713
-
Viscosity baths
41103714
-
Tissue flotation baths
41103715
-
Laboratory bath accessories or supplies
41103716
-
Blood bank plasma thawing bath
41103717
-
Tissue culture bath
41103718
-
Histology tissue freezing bath
41103801
-
Laboratory mixers
41103802
-
Roller mixers
41103803
-
Stirring tables
41103804
-
Multi bank or flocculation equipment
41103805
-
Laboratory vibrators
41103806
-
Magnetic stirrers
41103807
-
Laboratory touch mixers
41103808
-
Platelet mixers
41103809
-
Hematology or chemistry mixers
41103810
-
Overhead stirrers
41103811
-
Orbital shakers
41103812
-
Reciprocal shakers
41103813
-
Rotating shakers
41103814
-
Vortex mixers
41103815
-
Tube rotators
41103816
-
Mixer or shaker accessories or attachments
41103817
-
Laboratory reactor
41103818
-
Microplate shaker
41103819
-
Tissue culture roller drum
41103901
-
Microcentrifuges
41103902
-
Refrigerated microcentrifuges
41103903
-
Benchtop centrifuges
41103904
-
Refrigerated benchtop centrifuges
41103905
-
Floor centrifuges
41103906
-
Refrigerated floor centrifuges
41103907
-
Ultracentrifuges
41103908
-
Vacuum centrifuges
41103909
-
Centrifuge rotors
41103910
-
Centrifuge buckets
41103911
-
Centrifuge adapters
41103912
-
Centrifuge brushes
41103913
-
Laboratory centrifuge accessories
41103914
-
Cytocentrifuge
41103915
-
Cytocentrifuge cytofunnel or filter card or clamp
41103916
-
Centrifuge control board or printed circuit board
41103917
-
Centrifuge microplate carrier or sealing lid
41104001
-
Sample changers
41104002
-
Sample oxidizer
41104003
-
Sample preparation line
41104004
-
Sample preparation bombs
41104005
-
Laboratory bailers
41104006
-
Coliwasas
41104007
-
Water samplers
41104008
-
Air samplers or collectors
41104009
-
Air sampling pumps
41104010
-
Reagent kits for use with air samplers
41104011
-
Filters or other spare parts for samplers
41104012
-
Dust fall holders or jars
41104013
-
Sulphur dioxide or smoke samplers
41104014
-
Sample applicators
41104015
-
Plant samples analysis equipment
41104017
-
Sample holders
41104018
-
Solid phase extraction preparations
41104020
-
Calcine element flow tray
41104021
-
Fraction collector
41104022
-
Sample shaper
41104023
-
Water sampler accessories
41104101
-
Slide or specimen mailers or shippers
41104102
-
Lancets
41104103
-
Heel warmers
41104104
-
Tourniquets
41104105
-
Specimen collection or transport bags
41104106
-
Phlebotomy trays or accessories
41104107
-
Vacuum blood collection tubes or containers
41104108
-
Non vacuum blood collection tubes or containers
41104109
-
Blood unit collection bags
41104110
-
Blood culture bottles
41104111
-
Cytology collection kits or containers
41104112
-
Urine collection containers
41104114
-
Frepp Sepp collection containers
41104115
-
Serum Filter collection containers
41104116
-
Swab collection or transport containers
41104117
-
Specimen holders
41104118
-
Specimen collection container
41104119
-
Bone tissue collection containers
41104120
-
Sedimentation rate tubes
41104121
-
Stool collection containers with media
41104122
-
Stool collection containers without media
41104123
-
Sputum collection apparatus or containers
41104124
-
Laboratory bone marrow biopsy trays
41104125
-
Histology or pathology preservative specimen container
41104126
-
Histology or pathology specimen container
41104127
-
Laboratory feeder
41104128
-
Culture specimen collector without swabs
41104129
-
Neonatal metabolic disorder screen collection card
41104130
-
Blood culture collection kit
41104131
-
Pinworm collection paddle or device
41104132
-
Urine strainer for renal calculi
41104133
-
Umbilical cord blood collector
41104134
-
Refrigerant pack for diagnostic specimen shippers
41104135
-
Template bleeding time incision device and blotting paper
41104136
-
Specimen drop box
41104137
-
Arterial and capillary blood gas collection kit
41104138
-
Blood collection needle holders
41104201
-
Water purification reagents
41104202
-
Deionization or demineralization equipment
41104203
-
Base exchange equipment
41104204
-
Reverse osmosis equipment
41104205
-
Ultra violet water purification units
41104206
-
Ultra pure water systems
41104207
-
Water analysis systems
41104208
-
Dehydrators
41104209
-
Deoxiders
41104210
-
Dissolvers
41104211
-
Softeners
41104212
-
Water filtration cartridges
41104213
-
Distilled or deionized water
41104214
-
Reverse osmosis equipment parts and accessories
41104301
-
Standard fermentation units
41104302
-
Continuous culture apparatus
41104303
-
Anaerobic jars or accessories
41104304
-
Digestion systems
41104305
-
Inspissators
41104306
-
In vitro culture equipment
41104307
-
Microbiology fermentation equipment
41104308
-
Anaerobic environmental culture systems or supplies
41104309
-
Microaerophilic environmental cutlure system
41104401
-
Gravity convection general purpose incubators
41104402
-
Forced air or mechanical convection general purpose incubators
41104403
-
Tissue culture incubators
41104404
-
Cooled biological oxygen demand BOD incubators
41104405
-
Shaking incubators
41104406
-
Plate incubators
41104413
-
Dry wall single chamber carbon dioxide incubators with humidity control
41104414
-
Dry wall dual chamber carbon dioxide incubators with humidity control
41104415
-
Water jacketed single chamber three gas incubators
41104416
-
Water jacketed dual chamber three gas incubators
41104417
-
Water jacketed single chamber three gas incubators with humidity control
41104418
-
Water jacketed dual chamber three gas incubators with humidity control
41104419
-
Dry wall single chamber three gas incubators
41104420
-
Dry wall dual chamber three gas incubators
41104423
-
Refrigerated incubators
41104424
-
Incubator accessories
41104425
-
Carbon dioxide incubator
41104426
-
Media preparation apparatus
41104427
-
Platelet incubator
41104501
-
Laboratory mechanical convection ovens
41104502
-
Gravity convection ovens
41104503
-
Ageing ovens
41104504
-
Cleanroom ovens
41104505
-
Laboratory quartz oven pots
41104506
-
Laboratory safety ovens
41104507
-
Laboratory microwave ovens
41104508
-
Induction dryers
41104509
-
Vacuum ovens
41104510
-
Drying cabinets or ovens
41104511
-
Hybridization ovens or incubators
41104512
-
Laboratory oven accessories
41104513
-
Magnesium melt oven
41104601
-
Laboratory box furnaces
41104602
-
Programmable box furnaces
41104603
-
Tube furnaces
41104604
-
Programmable tube furnaces
41104605
-
Crucible furnaces
41104609
-
Laboratory safety furnaces
41104610
-
Laboratory furnace hearthplates
41104611
-
Laboratory furnace replacement insulation
41104612
-
Laboratory furnace accessories
41104613
-
Laboratory hot press
41104701
-
Freeze dryers or lyopholizers
41104702
-
Freeze dryer glassware
41104703
-
Tray dryers
41104704
-
Freeze dryer or lyopholizer accessories
41104801
-
Flask or retort units
41104802
-
Bi distillation units
41104803
-
Laboratory evaporators
41104804
-
Vacuum or rotary evaporators
41104805
-
Nitrogen blowdown evaporators
41104806
-
Extracting equipment for laboratories
41104810
-
Fractionation apparatus
41104811
-
Density gradient fractionators
41104812
-
Distillation pipings or columns or fittings
41104813
-
Reflux components
41104814
-
Laboratory heat exchange condensers
41104815
-
Kjeldahl nitrogen determination apparatus
41104816
-
Vacuum or centrifugal concentrators
41104817
-
Extraction thimbles
41104818
-
Structured packing
41104819
-
Cell harvester
41104820
-
Laboratory fractional distillation apparatus
41104821
-
Static sample concentrator
41104901
-
Laboratory line filters
41104902
-
Gel filtration equipment
41104903
-
Ultra filtration equipment
41104904
-
Sintered cell filters
41104905
-
Thin channel filtration equipment
41104907
-
Molecular filtration equipment
41104908
-
Laboratory cartridge element filters
41104909
-
Laboratory filter holders or cyclones
41104910
-
Laboratory multi sheet or press filters
41104911
-
Laboratory air filtration systems
41104912
-
Fluid presses filter
41104913
-
Bioseparation filters
41104914
-
Bottletops or filtration cups
41104915
-
Capsules filters
41104916
-
Centrifugal filters
41104917
-
Laboratory environmental filters
41104918
-
Laboratory glass filters
41104919
-
Laboratory HEPA filters
41104920
-
Hybridization filters
41104921
-
Laboratory membrane filters
41104922
-
Syringe filters
41104923
-
Multiwell plate filters
41104924
-
Microbiology filters
41104925
-
Laboratory filtration hardware or accessories
41104927
-
Filter support screens
41104928
-
Laboratory bottle receiver
41104929
-
Laboratory filter papers
41104930
-
Molecular sieve
41104931
-
Serum separator tube filter
41104932
-
Cell strainer
41104933
-
PCR System
41105001
-
Laboratory separators
41105002
-
Laboratory sifting equipment
41105003
-
Test sieves
41105101
-
Laboratory vacuum pumps
41105102
-
Peristaltic pumps
41105103
-
Laboratory centrifugal pumps
41105104
-
Syringe pumps
41105105
-
Metering pumps
41105106
-
Chromatography pumps
41105107
-
Laboratory drum pumps
41105108
-
Laboratory general purpose tubing
41105109
-
Rotary vane pumps
41105202
-
Hematology slide stainers
41105203
-
Microbiology slide stainers
41105204
-
Laboratory slide stainer accessories
41105205
-
Microslide making equipment
41105206
-
Immunohistochemistry autostainer
41105207
-
Immunohistochemistry autostainer accessory
41105208
-
Laboratory staining rack and tray
41105209
-
Multidepartment manual slide stainer set
41105301
-
Gel boxes
41105302
-
Gel dryers
41105303
-
Electrophoresis system power supplies
41105304
-
Transilluminators
41105305
-
Electrophoresis system accessories
41105307
-
Instrumentation for capillary electrophoresis
41105308
-
Capillaries or cartridges
41105309
-
Kits or reagents for capillary electrophoresis
41105310
-
Blotting or transfer accessories
41105311
-
Blotting or transfer apparatus
41105312
-
Combs or plates or spacers or trays
41105313
-
Cassettes or related detection accessories
41105314
-
Gel documentation systems
41105315
-
Gel documentation accessories
41105316
-
Ultraviolet crosslinkers
41105317
-
Agarose gel making reagents
41105319
-
Polyacrylamide gel making reagents
41105320
-
Polyacrylamide premade gels
41105321
-
Nucleic acid gels stain
41105322
-
Polyacrylamide gels stain
41105323
-
Electrophoresis premade buffers or solutions
41105325
-
Microwells plates for deoxyribonucleic acid DNA or deoxyribonucleic acid DNA hybridization
41105326
-
Hybridization reagents or buffers
41105327
-
Conjugated nucleotides or oligomers
41105328
-
Premade northern or southern or western blots
41105329
-
Blocking agents
41105330
-
Control proteins or cell lysates or tissue lysates
41105331
-
Protein chemifluorescent detection reagents or kits or substrates
41105332
-
Protein chemiluminescent detection reagents or kits or substrates
41105333
-
Protein chromogenic detection reagents or kits or substrates
41105334
-
Deoxyribonucleic acid DNA quantitation markers
41105335
-
Deoxyribonucleic acid DNA size markers or standards
41105336
-
Isoelectric focusing IEF markers
41105337
-
Protein electrophoresis markers
41105338
-
Ribonucleic acid RNA markers or standards
41105339
-
Blotting membranes
41105340
-
Electrophoresis system
41105341
-
Autoradiography film
41105342
-
Kit and reagent for agarose gel electrophoresis
41105343
-
Kit and reagent for tape based electrophoresis
41105344
-
Electrophoresis sample applicator or blade
41105345
-
Protein gel stain
41105501
-
Deoxyribonucleic acid DNA cleanup or gel extraction kits
41105502
-
Kits for deoxyribonucleic acid DNA extraction from food
41105504
-
Genomic deoxyribonucleic acid DNA purification kits
41105506
-
Kits for purification of messenger ribonucleic acid mRNA
41105508
-
Nucleic acids coprecipitants
41105509
-
Nucleic acids quantitation kits
41105510
-
Phage deoxyribonucleic acid DNA purification kits
41105511
-
Kits for plasmids deoxyribonucleic acid DNA extraction from yeast
41105512
-
Plasmids or cosmids or bacterial artificial chromosomes BAC purification kit
41105513
-
Labeled nucleic acid purification kits
41105514
-
Reagents for nucleic acid extraction or precipitation or resuspension
41105515
-
Ribonucleic acid RNA cleanup or stabilization materials
41105516
-
Ribonucleic acid RNA gel extraction kits
41105517
-
Kits for nucleic acid extraction from plant cells or tissue
41105518
-
Total ribonucleic acid RNA purification kits
41105519
-
Viral deoxyribonucleic acid DNA purification kits
41105520
-
Viral ribonucleic acid RNA purification kits
41105521
-
Deoxyribonucleic acid DNA detection system
41105601
-
Kits or enzymes for sequencing
41105700
-
Gene arrays
41105803
-
Transcription or translation systems or kits
41105804
-
Translation labeling accessories
41105902
-
Complementary deoxyribonucleic acid cDNA libraries
41105903
-
Complementary deoxyribonucleic acid cDNA synthesis kits
41105906
-
Protein or peptide display libraries
41105907
-
Two hybrid libraries or systems
41105908
-
Viral packaging kits
41106004
-
Nucleic acid non radioactive labeling kits
41106005
-
Nucleic acid radioactive labeling kits
41106101
-
Cytogenetics kits
41106102
-
Differential display or subtraction kits
41106103
-
Deoxyribonucleic acid DNA typing kits
41106104
-
Nuclease protection assays
41106203
-
Bacteria transformation kits
41106204
-
Bottled agar media or stabs for bacteria
41106210
-
Media or supplements for schizosaccharomyces pombe
41106211
-
Media ingredients or additives schizosaccharomyces pombe
41106212
-
Media ingredients or additives for bacteria
41106213
-
Premixed media dry
41106214
-
Reagents for preparing competent bacteria
41106215
-
Reagents for preparing competent yeast
41106216
-
Rich media for yeast
41106217
-
Specialty plates for bacteria
41106218
-
Specialty premixed media dry
41106219
-
Synthetic complete supplement mixtures for yeast
41106220
-
Synthetic media for yeast
41106221
-
Yeast competent cells
41106222
-
Yeast transformation kits
41106223
-
Yeast nitrogen bases YNB or yeast nitrogen base YNB variants
41106224
-
Electroporation system
41106232
-
Automated microbial culture plate streaker
41106302
-
Gene specific polymerase chain reaction PCR kits
41106303
-
Kits for polymerase chain reaction PCR purification
41106306
-
Polymerase chain reaction PCR buffers
41106307
-
Polymerase chain reaction PCR optimizing products
41106308
-
Polymerase chain reaction PCR or reverse transcriptase polymerase chain reaction RT PCR primers
41106309
-
Premade complementary deoxyribonucleic acid cDNA
41106311
-
Purified ribonucleic acids RNA
41106313
-
Reverse transcriptase polymerase chain reaction RT PCR kits
41106314
-
Thermostable deoxyribonucleic acid DNA polymerases or kits
41106315
-
Near patient PCR machine
41106401
-
Adaptors or linkers
41106402
-
Miscellaneous primers
41106403
-
Sequencing primers
41106501
-
Bacterial expression kits
41106502
-
Eucariotic transfection reagents
41106505
-
Insect expression kits
41106506
-
Insect media
41106507
-
Insect medium supplements or reagents
41106508
-
Mammalian cell expression kits
41106509
-
Mammalian cells
41106510
-
Kits for protein extraction from mammalian cells or tissues
41106511
-
Kits for protein extraction from bacteria
41106512
-
Kits for protein extraction from yeast
41106513
-
Reporter gene assay
41106514
-
Stable mammalian cell lines
41106515
-
Yeast expression kits
41106516
-
Enzyme expression consumables
41106601
-
Chromosome targeting vectors
41106604
-
Display vector maps or sequences
41106605
-
Enzyme reporter vector maps or sequences
41106606
-
Expression complementary deoxyribonucleic acid cDNA vectors
41106607
-
Fluorescent protein vector maps or sequences
41106608
-
Fusion vectors
41106609
-
Gene targeting vectors
41106610
-
General cloning vectors
41106611
-
Hybrid system vectors or kits
41106612
-
Insect expression vectors
41106615
-
Polymerase chain reaction PCR cloning vectors or kits
41106616
-
Phage or viral deoxyribonucleic acids DNA
41106617
-
Plasmid mutagenesis vectors or kits
41106619
-
Sequencing vectors
41106620
-
Signal transduction reporter vector maps or sequences
41106621
-
Virus mediated expression vectors or kits
41106622
-
Yeast expression vectors
41151629
-
Lab screening test kit
41151500
-
Clinical laboratory and toxicology testing systems, components, and supplies
41171600
-
Microbiology devices
New clarification added: Q76: As per Section III Returnable Bidding Forms, Form D: Technical Quotation Form, LOT 25: Semi-Automated Microtome. Please clarify, do you need a motorized microtome or, as in the name, a semi-motorized microtome? If semi-motorized, then you need to remove the Lines that refer to the motorized microtome: Lines 11-16, 36. Line 26 indicates Programmable sample retraction in the range not less than: 0 - 190 µm. This range is artificial and only 1 manufacturer has it. Please reduce the range to 0-50 µm, which is actually used in histological practice. As written earlier in other queries, Lines 22-25, 27 and 29-33 reflect parameters 1 of a single manufacturer. For other manufacturers, the trimming range does not exceed 600 microns, and besides, the ranges and step accuracy differ from the declared ones. Do we understand correctly that they would be considered appropriate microtomes with excellent ranges and minimal adjustment steps? For example, for Lines 22-25; 0.25~2.5 µm in 0.25 µm steps; 2.5~5.0 µm in 0.5 µm steps;5.0~10 µm in 1 µm steps10~30 µm in 2 µm steps;30~60 µm in 5 µm increments;60~100 µm in 10 µm steps;The question has already been asked about the slice transfer system. If we take into account all these parameters, then only 1 system meets the requirements. And such a system refers to scientific instruments. In the daily practice of a histologist, taking into account the number of sections produced, transfer to a water bath is carried out with a brush. Therefore, you need to decide whether you plan to purchase equipment from a specific manufacturer, or still on a competitive basis, a microtome and water bath system. If the latter, then it is necessary to remove paragraphs 45, 51, 52, 53. The control of the water bath in the vast majority of cases is carried out through the control panel with a display, usually monochrome. Request in Line 54 to remove color and add control through a control panel with a display or a touch screen. A76: Please consider the following for Lot 25: Semi-Automated Microtome: Line 11 will be rephrased into: ‘Optional Motorized operating modes’; Line 12 will be rephrased into: ‘Optional Continuous Mode’; Line 13 will be rephrased into: ‘Optional Program Mode’; Line 14 will be rephrased into: ‘Optional Single Step Mode’; Line 15 will be rephrased into: ‘Optional Gradual Step-by-Step Mode’; Line 16 will be rephrased into: ‘Optional Cutting Window Mode’; Line 36 will be rephrased into: ‘Optional adjustable cutting speed in the range not less than: 5 - 400 μm/sec’; Line 26 will be rephrased into: ‘Programmable sample retraction in the range not less than: 0 - 50 µm.';Lines 22-33 will be evaluated as compliant as per changes previously approved in Revision III. Line 45 will be rephrased into: ‘Optional - Automatic system for moving sections into a water bath, compatible with the microtome and allowing you to install the laminar flow of liquid or adequate solution’; Line 51 will be rephrased into: ‘Optional - The volume of water in the pump tank is not less than: 500 ml‘; Line 52 will be rephrased into: ‘Optional - Laminar fluid flow rate in the range not less than: 0 - 500 ml/min’; Line 53 will be rephrased into: ‘Optional - Adjustable "bridge" for waste cutting’; Line 54 will be rephrased into: ‘Control via control panel with a display or touch screen’.
Edited on:
24-Jun-2023 08:54
Edited by:
webservice@unops.org
New clarification added: Q75: As per Section III Returnable Bidding Forms, Form D: Technical Quotation Form, LOT 24: Slide Dryer: A slide drying system is planned for purchase. However, most of the requirements are for a built-in timer (Lines 7, 14-17), which looks odd and leads to one specific manufacturer. Can a non-built-in timer be suggested? A75: Please consider the following for Lot 24: Slide Dryer, please consider the following: Lines 7, 14-17 refer to any timer solution. A non-built-in timer will be evaluated as compliant, therefore Line 14 will be rephrased into:‘Optional continuous operation mode’; Line 15 will be rephrased into:‘Optional built-in real-time clock or adequate solution:’
Edited on:
24-Jun-2023 08:54
Edited by:
webservice@unops.org
New clarification added: Q74: As per Section III Returnable Bidding Forms, Form D: Technical Quotation Form, LOT 23: Paraffin Embedding Station: Line 6: shows the tweezer holders, while the kit does not indicate the presence of electric tweezers and, as a rule, most models do not have them. In addition, in practice, ordinary picents are used. In this regard, we ask you to remove Line 11. A74: Please consider the following for LOT 23: Paraffin Embedding Station, Line 6 is referring to ‘Temperature adjustment step no more than 1°C’; However, Line 11 refers to ‘Connection of electric heated tweezers:’ Line 11 will be rephrased into: 'Optional connection of electric heated tweezers'.
Edited on:
24-Jun-2023 08:53
Edited by:
webservice@unops.org
New clarification added: Q73: As per Section III Returnable Bidding Forms, Form D: Technical Quotation Form, LOT 22: Automated Tissue Processor. Line 5. Does the number of wax containers count with the wax drain container or just the storage containers? If only storage containers, please reduce to 3 paraffin containers. Line 17: Do we understand correctly that the vacuum settings should be in the specified range? A73: Please consider the following for LOT 22: Automated Tissue Processor, Line 5 assumes wax drain containers, therefore remain unchanged; Line 17 is correctly understood, vacuum settings should be in the specified range; However, Line 17 will be rephrased into: ‘Vacuum adjustment in the range not less than: 0.4 - 1.2 bar or adequate’.
Edited on:
24-Jun-2023 08:52
Edited by:
webservice@unops.org
New clarification added: Q72: As per Section III Returnable Bidding Forms, Form D: Technical Quotation Form, LOT 19:Cryostat and reagent (frozen section): Line 5. Please expand on options for external camera illumination. Line 7: Only 1 manufacturer has two Peltier elements. All the rest have 1 Peltier element as standard. Line 16: Extend the horizontal feed range to 22mm. Please make changes to the competitive procedure. A72: Please consider the following for LOT 19: Cryostat and reagent (frozen section), Line 5 will be rephrased into:’Camera Interior or Exterior Lighting’; Line 7 will be rephrased into: ‘Stations for quick freezing based on the Peltier element at least: 1’; Line 16 will be rephrased into: ‘Horizontal feed not less than 20 mm ±20%’.
Edited on:
24-Jun-2023 08:52
Edited by:
webservice@unops.org
New clarification added: Q71: Reference is made to the RFQ-2023-45672_Section_III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV3_21June23.xlsLot 25 Semi-Automated Microtome. Line 36:Adjustable cutting speed in the range not less than 5 - 400 μm/sec. This parameter limits competition and excludes some Semi-automatic Rotary Microtome manufacturers from the competition. Usually, this characteristic applies for Fully-automatic Rotary Microtomes. Please consider changing this requirement in to ‘Optional Adjustable cutting speed in the range not less than 5 - 400 μm/sec’. A71: Please consider the following for Lot 25 Semi-Automated Microtome: Line 36 will be rephrased into:‘Optional adjustable cutting speed in the range not less than 5 - 400 μm/sec’.
Edited on:
24-Jun-2023 08:52
Edited by:
webservice@unops.org
New clarification added: Q70: Is it allowed to submit documents in Russian, or Uzbek or all documents must be translated into English? What documents are allowed to be submitted in Russian, or Uzbek? A70: Language of the quotation must be English. Bidders must provide a user manual with each item, in English and Russian Languages. If documentation in Russian is not available at the time of bidding, bidders shall confirm in the relevant returnable bidding form that they shall provide the requested documentation in Russian together with the equipment.
Edited on:
24-Jun-2023 08:50
Edited by:
webservice@unops.org
New amendment added #6: The RFQ is further amended as follows:To reflect the changes that address clarifications on various lines of Lot 6, Lot 13, Lot 18, Lot 19, Lot 22, Lot 23, Lot 24, Lot 25 and Lot 29, the document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV3_21June23" is replaced with the updated document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV4_23June23".Deadline for the submission of quotations is extended for 7 more days until 3 July 2023.All other unchanged terms and conditions of the RFQ remain in full force.
Edited on:
24-Jun-2023 08:49
Edited by:
webservice@unops.org
New clarification added: Q69: Reference made to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, LOT 6: Flow Cytometer and Reagents; Please accept the following clarifications for adequate competition. More powerful lasers can be used for accurate cell and fluorescence detection. Please use the following wording for Lines 15-17, Line 15: Blue source wavelength, power: 488 - 493 nm, 18mW or more; Line 16: Red source wavelength, power: 633 - 640 nm, 40mW or more; Line 17: Violet source wavelength, power: 405 - 407 nm, 40mW or more; Modern flow cytometers often use detectors other than PMT. For example, APD detectors. They have higher sensitivity and improved quantum efficiency with respect to PMT. Please change the interpretation. Line 24:Type of fluorescence detectors: Photomultiplier detector (PMT) or equivalent (APD and etc.), Line 31: Minimum sheath flow rate: 5 ml/min or better”. Do we understand correctly that if the sheath flow rate values are 5 ml/min or higher, then the parameter will be appropriate? Lines 20 and 21: indicate, respectively, the minimum number of detectors 10. i.e. the minimum number of registered parameters must be at least 10. In this case, Line 36 conflicts and limits competition, because it lists at least 20 parameters, which currently exist only for scientific devices, and if we understand correctly, a medical device is needed. For medical cytometers, the maximum number of parameters varies from 14 to 16, but this is already taking into account the maximum configurations. Please state in the following edition Line 36 Data acquisition rate of minimum 30 000 events per second with the acquisition of at least 10 parameters. Line 38: Please write "cleaning and/or decontamination fluids" as a different principle of cleaning and decontamination can be used, as well as the ability to connect different liquids to one compartment. Line 41 What kind of biohazardous samples are you planning to use? Line 42 please expand decontamination options The fluidics reservoir must be autoclavable or have other decontamination options and preferably equipped with pressure and vacuum readout. The flow cytometer manual sample feed system does not support 15 ml centrifuge tubes. These tubes can be used in cell sorters for sample collection, but this is a completely different device and purpose of use other than indicated. Maybe you meant 1.5 ml microcentrifuge tubes. Please make a change. Line 44:The manual sample feeding system should be able to hold different types of tubes, minimum but not limited to: 2 ml, 5ml (12 x 75 mm). Line 46:Built-in sample temperature control. As a rule, such a requirement is typical for cell sorters or some scientific cytometers. For medical cytometers that are used in diagnostics, there is no sample temperature control. The temperature control of the instrument, detectors and lasers is mandatory to check the stability of the instrument and avoid overheating, but not the control of the sample temperature. Please remove this item, or change it to control the temperature of the device. Line 47: shows only two types of automatic sample mixing (vortexing or shaking). Please expand this to just automatic sample mixing: Line 47: Automatic vortexing or shaking of tubes and plates or automatic sample mixing. Add English as an alternative for Line 50. A69: Please consider the following for LOT 6: Flow Cytometer and Reagents: lines 15, 16, and 17 remain unchanged. Bidders are encouraged to offer technical solutions better than required since requirements are stipulated as minimal. Therefore all offers above stipulated technical requirements will be evaluated as compliant. Line 24 will be rephrased into: ‘Type of fluorescence detectors: Photomultiplier detector (PMT) or equivalent’. Please refer to the response for lines 15-17; Line 31 will be evaluated as compliant if the values offered are 5 ml/min or higher. Line 36 will be rephrased into: ‘Data acquisition rate of minimum 30 000 events per second with the acquisition of at least 10 parameters’. Line 38 will be rephrased into: ‘External stand or inbuilt compartment for placing bottles of fluidic (sheath) solution, cleaning and/or decontamination fluids and waste containers in volumes specific for the offered equipment.’. There will be no biohazard samples in routine clinical procedures, therefore Line 41 will be rephrased into: 'Optional bio-hazard containment facility for probe washing' Line 42 will be rephrased into: ‘The fluidics reservoir must be autoclavable or similar decontamination option, preferably equipped with pressure and vacuum readout. Line 43 will be rephrased into: ‘Optional automatic sample loader to accommodate a minimum of 20 tubes and at least one microplate 96-well (U, flat or V-shaped bottom)’. Line 44 will be rephrased into: ‘The manual sample feeding system should be able to hold different types of tubes, minimum but not limited to 2 ml, 5 ml (12 x 75 mm) to 15 ml tubes or equivalent’. Line 46 will be rephrased into: ‘Optional system for temperature control’. Line 47 will be rephrased into: ‘Automatic sample mixing’. Line 50 will be rephrased into: ‘Software in English and/or Russian language’.
Edited on:
24-Jun-2023 08:48
Edited by:
webservice@unops.org
New clarification added: Q68: Reference is made to the RFQ-2023-45672_Section_III_Returnable Bidding Forms_Form D_Technical specifications for goods, Lot 23 Paraffin Embedding Station: Line 5 - Adjustable operating temperature - from room temperature to not less than 70°C.The operating temperature generally refers to the allowable ambient temperature. The adjustable temperature of a paraffin dispensing module and pre-warming module usually starts from 30°C or higher. For paraffin embedding, a melting temperature below 30°C is not required. Please specify the requirement. Line 11 - Connection of electric heated tweezers. Please specify the requirement. Does this requirement mean the possibility of connecting electric heated tweezers or the presence of this accessory? A68: Please consider the following for Lot 23, Paraffin Embedding Station: Line 5 will be rephrased into: 'Adjustable operating temperature in the range 30-70°C'. whereas Line 11 will be rephrased into: 'Optional connection of electrically heated tweezers'.
Edited on:
24-Jun-2023 08:46
Edited by:
webservice@unops.org
New clarification added: Q67: Reference is made to the RFQ-2023-45672_Section_III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV3_21June23.xls:Lot 19 Cryostat and reagent (frozen section): Line 22 Color touchscreen control: This parameter limits competition and excludes some manufacturers from the competition. The combination of display and control panel allows you to make all the necessary settings. The absence of a touch screen does not affect the technical and functional qualities of the equipment. We would like you to clarify the requirement whether color touchscreen control is mandatory or change the requirement to "Screen control or equivalent". A67: Please consider the following for LOT 19 Cryostat and reagent (frozen section): Line 22 will be adjusted into: 'Screen control or equivalent'.
Edited on:
24-Jun-2023 08:44
Edited by:
webservice@unops.org
New clarification added: Q66: Please increase the delivery time for lots 2,3,4,6,13,19,22,23,24,25,28 due to the high volume of goods supplied and the complexity of the equipment up to 90 working days. A66: The delivery time mentioned as 90 Days, Tashkent, Uzbekistan – DPU (Incoterms2020)
Edited on:
24-Jun-2023 08:42
Edited by:
webservice@unops.org
New clarification added: Q65: Reference made to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, LOT 18-Anatomic pathology with IHC (immunochemistry) lab consumables- will all the requested reagents need to be registered in case of awarding? A65: Upon signature of a contract with UNOPS, the successful supplier shall submit the application form(s) along with the required documents for registration to Uzpharm Control and SES and shall apply efforts to obtain the certificate(s) of registration or authorization (s) or waiver(s) for the import and use of medical devices prior to goods’ arrival to the country. Please note that per the decree of the President of the Republic of Uzbekistan #6221 d.d. 05 May 2021, Point #12 on the special process of registration of medical products certified by certain international organizations (see Appendix # 3 to the decree for the full list of organizations), according to which products with CE or FDA registration can be registered in Uzbekistan within 15 working days. https://lex.uz/ru/docs/5411138#5419395.
Edited on:
24-Jun-2023 08:40
Edited by:
webservice@unops.org
New clarification added: Q64. Reference made to Section II: Schedule of requirement, an ISO 9001 certificate is not a mandatory requirement for doing business in the Republic of Uzbekistan. The mandatory requirement for an ISO 9001 certification certificate or a similar document is a restriction of competition. Please reconsider this requirement and allow non-ISO 9001-certified companies to participate in this tender A64. As per Section II.2b - Manufacturer & Supplier Certification, Requirements for Distributors, Quality Management System (preferably compliant to ISO 9001 or other applicable quality management system requirements). It is clearly stated that any QMS system will be accepted, although ISO 9001 is preferable, according to international standards. Therefore any QMS, including local certificates issued in Uzbekistan will be evaluated as compliant.
Edited on:
24-Jun-2023 08:31
Edited by:
webservice@unops.org
New clarification added: Q63: As per Section III Returnable Bidding Forms, Form D Technical Quotation Form, Lot 13-PCR Analyzer, real-time: Line 6:Temperature regime (operating) in the range not less than 300-990˚C.The temperature regime (operating) of the Thermocyclers is indicated incorrectly. The standard temperature regime of Thermocyclers is indicated in the range of 30-100˚C. Line 4:The maximum volume of the sample and the reaction mixture shall not exceed 150 ml.The volume of the reaction PCR mixture is incorrect. The standard volume of the PCR reaction mixture is 1-50 µl (microliters), max volume of 150 µl (microliters).Line 25:Forester Resonance Power Transfer (FRET) function. The correct interpretation of the term FRET - is fluorescence resonance energy transfer. A63: Please consider the following for Lot 13- PCR Analyzer, real-time, Line 6 will be adjusted into: 'Temperature regime (operating) in the range not less than 30-99˚C'.; Line 4 will be adjusted into: 'The maximum volume of the sample and the reaction mixture shall not exceed 150 µl (microliters)' ; Line 25 will be adjusted into: 'Fluorescence Resonance Energy Transfer (FRET) function'.
Edited on:
24-Jun-2023 08:27
Edited by:
webservice@unops.org
New clarification added: Q62: Reference made to Evaluation criteria, Item "Experience: The Bidder shall be in the continuous business of manufacturing and/or supplying the products as specified in the ‘Section II: Schedule of requirements’ during the last two (2) years OR similar products during the last three (3) or more years prior to bid opening." What documents shall a Bidder submit? Is a certificate of registration (indicating the date when a Bidder sharted its business activity) enough to meet this criterion? A62: No, to meet this requirement a certificate of registration is insufficient. You also need to mention the relevant experience in Form E - Experience Form.
Edited on:
24-Jun-2023 08:11
Edited by:
webservice@unops.org
New clarification added: Q61: Answer 36 is not in line with the answer we received previously (A 14), Could you please confirm if we need to offer the requested set of reagents for each machine (for example,17 units in case of LOT 3) or if it is a single order for reagents? Moreover for lot 3, does this apply to all the reagents (from lines 50 to 74)? We would kindly ask to receive a second extension of the deadline so that we can revise the specifications with the manufacturers, according to the latest amendment. A61: As per previous responses, all components, spare parts and consumables are stipulated per item, therefore they are specific for each instrument. However, general rules and requirements also apply, as per Sections II and III of the RFQ/2023/45672, which means that each device must be delivered with consumables in quantities specified per item. For lot 3 it means all 17 items must be delivered with 17 packages of requested consumables, lines 49-75.
Edited on:
24-Jun-2023 08:10
Edited by:
webservice@unops.org
New clarification added: Q60: As per Section III Returnable Bidding Forms, Form D: Technical Quotation Form, Lot 29-Refrigerated centrifuge, line 26: Tubes installed in the rotor: 90 ml not less than 200 pcs. Please recheck the volume of the requested test tubes as 90 ml tubes are not produced or produced very rarely and tubes of this volume are not listed in any methods in accordance with the use of this equipment. A60: Please consider the following for Lot 29, Refrigerated centrifuge: Line 26 will be rephrased into 'Tubes installed in the rotor: 90 ml or similar, not less than 200 pcs'.
Edited on:
24-Jun-2023 08:09
Edited by:
webservice@unops.org
New clarification added: Q58: As per Section III Returnable Bidding Forms Form D Technical Quotation Form, Lot 29, Refrigerated centrifuge, please consider line 8:The volume of the working chamber is not less than 80 l, and line 9 Centrifuges with this chamber volume for a maximum volume of 620 ml are inconsistent figures, please make adjustments if appropriate. A58: Line 8 remains unchanged: 'The volume of the working chamber is not less than 80 l', whereas Line 9: will be rephrased into 'The maximum capacity not less than 4x600ml'.
Edited on:
24-Jun-2023 08:07
Edited by:
webservice@unops.org
New clarification added: Q59: As per Section III Returnable Bidding Forms Form D Technical Quotation Form, Lot 29, Refrigerated centrifuge, line14: Operation of the rotor at ambient temperature from -5 to +600°C and relative humidity of the medium up to 80% at a temperature of +250°C. At such temperatures, the rotor may be deformed. Please clarify this requirement. A59: Please consider the following for Lot 29, Refrigerated centrifuge, line14 will be rephrased into: 'Operation of the rotor at ambient temperature from -5 to +60°C and relative humidity of the medium up to 80% at a temperature of +25°C:'
Edited on:
24-Jun-2023 08:05
Edited by:
webservice@unops.org
New clarification added: Q57: Where do we upload IFU, brochure, and other sources that we used for filling out of RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods_Comparative Data Table? There is no section or place that is identified to upload the relevant documents. A57: Please see the Checklist section of the RFQ. You can upload these documents under "Highlighted Copies of Original Brochures, Data Sheets, Drawings, Pictures, User and/or Service Manuals and Certificates with Related Test Reports Produced by the Manufacturer" section for each specific lot you're bidding for.
Edited on:
23-Jun-2023 07:07
Edited by:
webservice@unops.org
New amendment added #5: The RFQ is further amended as follows:Requirement on Quality Management System for Bidders is broadened. RFQ Sections II and III as well as Checklist are updated to reflect this change. To reflect the change described in point 1 above the document titled "RFQ-2023-45672_Section_II_Schedule_of_Requirements_Rev1_31May23" is replaced with the updated document titled "RFQ_2023_45672_Section_II_Schedule_of_Requirements_Rev2_21Jun23.To reflect the change described in point 1 above and address the requested clarifications on various lines of Lot 13, Lot 19, Lot 22, Lot 23 and Lot 25, the document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV2_12June23" is replaced with the updated document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV3_21June23".Deadline for clarification requests is extended until 23 June 2023.All other unchanged terms and conditions of the RFQ remain in full force.
Edited on:
21-Jun-2023 18:11
Edited by:
webservice@unops.org
New clarification added: Q56: Please see the below comments on technical requirements on LOT 23 Paraffin Embedding Station; Line 4:The volume of the paraffin container is not less than: 5,0 L .The limitation on the minimum size of a paraffin oven restricts fair competition. 4 liters of paraffin is enough to embedding a day's volume of work in the laboratory. Please specify a value of at least 4,0 L; Line 10 Built-in wax trimmer and paraffin drainage system in a separate container. This function can be implemented through the heating zone without a trimmer on any Paraffin Embedding Station. This value is intended to limit competition. Please delete this paragraph from the Specification.Line12:Number of compartments for heating tweezers not less than: 8. Please indicate the most common value (Number of compartments for heating tweezers not less than: 6) to not limit the competition. A56: Please consider the following for LOT 23 Paraffin Embedding Station: Line 4 will be adjusted into: ‘The volume of the paraffin container is not less than: 4,0 L.; Line 10 will be adjusted into a ‘Built-in wax trimmer and paraffin drainage system in a separate container or adequate solution:’ Line 12 will be adjusted into ‘Number of compartments for heating tweezers not less than: 6’.
Edited on:
21-Jun-2023 17:56
Edited by:
webservice@unops.org
New clarification added: Q55: Please see the below comments on technical requirements: LOT 22 - Automated Tissue Processor: Line 5: Number of containers for paraffin not less than: 5 pcs. This value is intended to limit competition. This parameter corresponds to only one model of tissue processors from a specific manufacturer. In this manufacturer's model, only 4 paraffin containers are functional, the fifth is not connected to the device and is used as a spare container. Please specify a value of at least 4 pcs. Line 6: Heating of containers for reagents from room temperature not less than: 60°C. This value is intended to limit competition. Please change this value to 50 degrees, since temperatures above 50 degrees during the dehydration stage can lead to the death of tissue samples and are not used in laboratories. Line 8: Special element for preheating of reagents. This parameter corresponds to only one model of tissue processors from a specific manufacturer. This value is intended to limit competition. Please remove this item from the requirements. Line 10: Fast processing time, not more than: 45 min. The run speed depends on the run protocol, reagents, and sample size. This value is intended to limit competition. Please remove this item from the requirements. Line 12: The number of processing protocols in memory is not less than: 18. Histological laboratories use several standard protocols (4-6) on a routine basis. This value is intended to limit competition. Please change the values to - The number of processing protocols in memory is not less than: 10. Line 19: Control via colour touch screen (at least 15" diagonal):This requirement does not make functional sense as the software adapts to the screen size according to UX/UI approaches. Please exclude requirement - at least 15" diagonal.Line 24: Built-in UPS with support for at least 4 hours. The built-in power sources in vacuum processors are not capable of ensuring the operation of the device for more than 5-10 minutes. A UPS capable of running a 1.5-2 kVA/h device would be about the size of a vacuum processor. This clause is intended to limit competition. Please remove this item from the requirements. A55: Please consider the following: for LOT 22- Automated Tissue Processor: Line 5 will be adjusted into: ‘Number of containers for paraffin not less than: 4 pcs.’; Line 6 will be adjusted into ‘Heating of containers for reagents from room temperature not less than: 50°C’; Line 8 will be adjusted into ‘Optional special element for preheating of reagents:’; Line 10 will be adjusted into ‘Optional fast processing time.’; Line 12 will be adjusted into ‘The number of processing protocols in memory is not less than: 10’; Line 19 will be adjusted into ‘Control via colour touch screen:’; Line 24 will be adjusted into ‘Built-in UPS’.
Edited on:
21-Jun-2023 17:45
Edited by:
webservice@unops.org
New clarification added: Q54: With reference to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, Lot 15: Microarray Scanner: are solid tumor specimens (FFPE) also of interest? and should both liquid and solid tumors be considered? A54: Samples of both solid tumors and liquid OCs and lymphomas are of interest.
Edited on:
21-Jun-2023 17:37
Edited by:
webservice@unops.org
New clarification added: Q53: With reference to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, Lot 15: Microarray Scanner: in Section A, 1. The General Description says: The system is designed to qualitatively and quantitatively determine the metastatic burden by measuring the expression of CK19 mRNA as a marker: If only one gene (CK19) needs to be tested, why are they looking for Microarray as a solution? A53: Section A1 - Qualitative and quantitative analysis of metastatic burden is required by measuring the expression of mRNA of various markers in oncopathologies.
Edited on:
21-Jun-2023 17:35
Edited by:
webservice@unops.org
New clarification added: Q52: With reference to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, Lot 15: Microarray Scanner: it is important to know if the end user wants to perform a gene expression analysis or if the analysis is needed to identify copy number gain and loss, LOH, cnLOH, clonal heterogeneity and ploidy status, and mosaicism. Or they may need to do both. If the intended use is gene expression analysis, then which genes are of interest? A52: It is necessary to analyze the expression of oncopathology candidate genes, clonal heterogeneity, and mosaicism.
Edited on:
21-Jun-2023 17:31
Edited by:
webservice@unops.org
New clarification added: Q51: With reference to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, Lot 15: Microarray Scanner: what should be the target of testing, since we need to know which analyses to choose? A51: To study the presence of mutations in candidate genes for oncopathologies of various genesis in patients of the Uzbek ethnic group, in order to compile a genetic passport, which is necessary for personalized therapy.
Edited on:
21-Jun-2023 17:29
Edited by:
webservice@unops.org
New clarification added: Q50: Referring to General Conditions of contract clause 7. Insurance and Liability – Will the Customer have an Employer's Liability Insurance policy or do they need some other liability insurance? (clause 7.2.4 - indicates that there may be other insurance agreed in writing with UNOPS.) A50: The Customer shall not have an Employer's Liability Insurance and shall not need any other liability insurance, provided that the Supplier (the awarded contractor) fulfils its insurance obligations per the Schedule of Requirements and these GCCs.
Edited on:
21-Jun-2023 17:26
Edited by:
webservice@unops.org
New clarification added: Q49: How will the goods be accepted and where, at our warehouse in Tashkent or at the Customer's warehouse with further delivery to end users? A49: Reference made to Section II - Schedule of requirements Rev2_ 21 June 2023, once goods are delivered to final destinations, the Supplier shall then arrange for and carry out the related services (installation, assembly, calibration (where necessary), testing and commissioning and end-user training activities) in each hospital for the acceptance of Goods.
Edited on:
21-Jun-2023 17:17
Edited by:
webservice@unops.org
New clarification added: Q48: Please see the below comments on technical requirements: LOT 19 Cryostat & Reagent - line 10 - The thickness of slices and alignment slices in the range is not less than: 1 - 300 μm. This indicator is not applicable in practice and is intended to limit competition. Trimming and cutting more than 40-80 µm can lead to the death of the sample or its lack for diagnosis. Please specify a value of at least 1 - 99 μm. Line 23 - Russian language support in the user interface. Purchase restriction, please allow English. A48: RFQ Section III - Returnable Bidding Forms, Form D, Lot 19, line 10 will be adjusted into: ‘The thickness of slices and alignment slices in the range is not less than: 1 - 99 μm; Line 23 will be adjusted into "English and/or Russian language support in the user interface".
Edited on:
21-Jun-2023 17:12
Edited by:
webservice@unops.org
New clarification added: Q47: What information should be indicated on the product labeling during shipment? - Paragraph 5.3 of the General Conditions states that this information should be provided by UNOPS. A47: Specific Packing and Shipping instructions shall be provided together with the Purchase Order upon award of contract.
Edited on:
21-Jun-2023 17:08
Edited by:
webservice@unops.org
New clarification added: Q46: Referring to General Conditions of contract, Clause 2. Responsibility for employees - which personnel does this requirement apply to? A46: This requirement applies to the Contractor’s officials, employees, agents, servants, subcontractors and other representatives (collectively, the Contractor’s “personnel”).
Edited on:
21-Jun-2023 17:05
Edited by:
webservice@unops.org
New clarification added: Q45: Referring to General Conditions of contract clause 5. Purchase of goods - what kind of shipping documents do we need to provide? Will it be at our discretion? Payment is also tied to the provision of documents, but it is not specified which documents are referred to (paragraph 8, Section I: Instructions to Bidders). Please clarify. A45: The complete list of the required documents for shipping as well as for payment purposes shall be specified in the Contract. Such documentation usually includes, without limitation, bills of lading, airway bills, certificate of origin, packing list, commercial invoice, authorisation for import, etc.
Edited on:
21-Jun-2023 17:02
Edited by:
webservice@unops.org
New clarification added: Q44: What are the payment terms? A44: Please read the Tender Particulars.
Edited on:
21-Jun-2023 16:53
Edited by:
webservice@unops.org
New clarification added: Q42: With reference to Form D: Technical Quotation Form, Lot 13 - Real-Time PCR Analyzer, Section D, lines 39 to 45, can you please provide the specs on this? For SNP (Single Nucleotide Polymorphisms) we need to know the end targets. A42: Study of polymorphism of genes involved in the detoxification of xenobiotics and present in oncopathologies of various genesis (breast cancer, CRC, cervical cancer, RTM, OC, RC, RL, etc.) in the Uzbek ethnic group.
Edited on:
21-Jun-2023 16:48
Edited by:
webservice@unops.org
New clarification added: Q41: Reference is made to Section III - Returnable Bidding Forms, Form D - Technical Quotation Form: Lot 13. 1) In line 6 the text says "Temperature regime (operating) in the range not less than: 300-990˚C". We believe that an extra zero is indicated in the specified required parameters, please adjust the parameters and indicate "30-99˚C". 2) In line 8 the text says "The accuracy of temperature maintenance is min: ± 0.30˚C". We believe that an extra zero is indicated in the specified required parameters, please bring the parameters in line and indicate "0.3˚C". 3) In line 9 the text says: "Maximum heating/cooling rate of the fusel not less than: 50˚C/sec". We believe that an extra zero is indicated in the indicated required parameters, please bring the parameters in line and indicate 5˚C / sec. A41: As per Section III Returnable Bidding Forms Form D Technical Quotation Form: Lot 13, please consider the following: (1) Line 6 will be amended into 'Temperature regime (operating) in the range not less than : 30-99˚C'; (2) Line 8 will be amended into 'The accuracy of temperature maintenance is min: ± 0.3˚C; (3) Line 9 will be amended into 'Maximum heating/cooling rate of the fusel not less than : 5˚C/sec'.
Edited on:
21-Jun-2023 16:43
Edited by:
webservice@unops.org
New clarification added: Q40: With reference to Form D: Technical Quotation Form, Lot 17, please clarify: - what is “Drug holder” listed in the requirements? - meaning of “Spectral karyotyping” - meaning of “Onchromosome scrubbing to track non-overlapping chromosome segments”? A40: Line 23 'Drug holder' refers to the holder of the sample/drug; Line 62 'Spectral Karyotyping' refers to Spectral karyotype (SKY) in which the homologous pairs of chromosomes are manipulated in such a way that they have distinctive colors. The SKY technique makes it easier to detect chromosomal abnormalities, as compared with a conventional karyotype. Line 81 'Onchromosome scrubbing to track non-overlapping chromosome segments' refers to the algorithm for segmentation and classification of touching and overlapping chromosomes in the metaphase images, which allows the extrication of overlapping chromosomes.
Edited on:
21-Jun-2023 16:25
Edited by:
webservice@unops.org
New clarification added: Q39: Can a local company (resident of the Republic of Uzbekistan) offer compatible goods, that are already cleared from Customs (by this company) and located in the Republic of Uzbekistan? A39: Yes.
Edited on:
21-Jun-2023 16:24
Edited by:
webservice@unops.org
New clarification added: Q38: Can you please clarify if, for LOTs 7-8-9-10, IQ/OQ protocol is requested for installation? A38: As per Section II.2d - Assembly and Calibration and Section II.2e - Online or On-Site Training Services, a standard completion report based on any pharmacopeia is required.
Edited on:
21-Jun-2023 16:20
Edited by:
webservice@unops.org
New clarification added: Q37: Regarding Section II: Schedule of Requirements (Rev1_of_31May23), Section II.1: Summary of the required goods and services, paragraph 2 «transportation from FCA point(s) to DPU Tashkent, Uzbekistan» and paragraph 4 «Custom clearance of the shipments shall be the responsibility of the Consignee». According to the Customs Code of the Republic of Uzbekistan, the resident company can only sell goods under procedure “import regime 40” (release for free circulation), which is analogous to the delivery under DDP Incoterms 2020 procedure. Therefore, a company resident of the Republic of Uzbekistan cannot deliver goods on FCA or DPU terms. Please clarify the possibility of delivery from a company resident of the Republic of Uzbekistan under DDP terms (Incoterms 2020). A37: If the goods that are already imported into the country, then resident firms shall offer them on FCA Tashkent terms. If goods are not yet delivered into the country, then resident firms can order the goods to be shipped directly on the name of the Consignee under DPU terms instead of first delivering the goods into the country on their own name and then handing them over to the Consignee under DDP terms.
Edited on:
21-Jun-2023 16:18
Edited by:
webservice@unops.org
New clarification added: Q36: With reference to Form D: Technical Quotation Form, - Lot 2: "All reagents and consumables required to perform at least 300 analyses (line No. 50)" - Lot 3: the number for reagents such as "A set of reagents for determining ALT activity: 1000 tests (line NO. 52)". Are these quantities for each machine or in total for all the 17 units? A36: Line 50 of Lot 2 and Line 52 of Lot 3 refer to the volumes of reagents for all 17 units, not for each machine. -
Edited on:
21-Jun-2023 16:10
Edited by:
webservice@unops.org
New clarification added: Q35: Reference is made to Section III_Returnable Bidding Forms_Form D_Technical specifications for goods_Comparative Data Table, Lot 4 "Automated immunoassay analyzer", line 5: Time to first result: no more than 30 minutes. But the preparation time of any reagent, taking into account incubation and reading exceeds 30 minutes. What is meant by: Time to first result no more than 30 minutes? A35: Line 5 refers to the analytical phase from putting the sample into the system to getting results for any of the assays listed in line 4.
Edited on:
21-Jun-2023 16:07
Edited by:
webservice@unops.org
New clarification added: Q.34: Regarding LOT 6 - Flow Cytometer and Reagents – 4 units, "certificate of registration for the import and use of the medical devices from the "Uzpharm Control". Currently, there are no reagents for flow cytometry, that are registered in "Uzpharm Control" in the Republic of Uzbekistan. Can you, please, clarify if there is an option to supply the reagents needed (in addition to the equipment) without the certificate of registration from the "Uzpharm Control"? A.34: Upon signature of a contract with UNOPS the successful supplier shall submit the application form(s) along with the required documents for registration to Uzpharm Control and SES and shall apply efforts to obtain the certificate(s) of registration or authorization (s) or waiver(s) for the import and use of medical devices prior to goods’ arrival to the country. Please note that per the decree of the President of the Republic of Uzbekistan #6221 d.d. 05 May 2021, Point #12 on the special process of registration of medical products certified by certain international organisations (see Appendix # 3 to the decree for the full list of organisations), according to which products with CE or FDA registration can be registered in Uzbekistan within 15 working days. https://lex.uz/ru/docs/5411138#5419395.
Edited on:
21-Jun-2023 16:06
Edited by:
webservice@unops.org
New clarification added: Q33: Could you please assist me with the following request which is mentioned in the Form B - Quotation submission form? "We have examined and have no reservations about the bidding documents, including amendments No.: [Insert the number and issuing date of each amendment]". Since we don't have an amendment which number we need to mention? A.33: This RFQ was amended 5 times, (1) on 31-May-23, (2) 07-Jun-23, (3) 08-Jun-23, (4) 12-Jun-23 and (5) 20-Jun-23. All amendments are published under the revisions tab of the RFQ. In the relevant section of the Form B bidders must insert the number and the issuing date of each amendment to confirm that they read them and considered them during their bid submission.
Edited on:
21-Jun-2023 16:03
Edited by:
webservice@unops.org
New clarification added: Q32: Some of the goods which we are planning to offer are already registered, but for some of them, there is no registration. Is it possible to send a letter of guarantee stating that the registration of this property will start immediately after the signing of the contract? A32": There is no requirement for a letter of guarantee during the bid submission. All you need to do is to tick the "YES" box in the Form D against the relevant service requirement.
Edited on:
21-Jun-2023 15:56
Edited by:
webservice@unops.org
New clarification added: Q31. If a bidder uses the services of a local subcontractor for post-delivery services (assembly, on-site training, calibration, services for hight-technology equipment), does this local company need to provide any information, other than the full legal name and address of subcontractors in "Form C: Price schedule form"? A31. In addition to providing the proposed subcontractor's name in Form C, the relevant experience of the proposed subcontractor must be mentioned in the Form E - Experience Form.
Edited on:
21-Jun-2023 15:53
Edited by:
webservice@unops.org
New clarification added: Q30: Reference is made to Form D, Lot 25 - Semi-Automated Microtome, Line 19: Although some modification has been implemented in the specifications, it is still unclear. A complete wheel turn means one section under any circumstances. No more than one section per wheel turn. So still this specification remains unclear. Line 43: This parameter limits competition and excludes most of the competing brands. It is not widely used in routine work and needs a thorough discussion to conclude that it displays a benefit for the histotechnicians. Line 45: Kindly clarify the need for this accessory. We have checked the list of hospitals where the microtome will be installed, and apart from the institution in Tashkent, the rest of facilities will have only one microtome for the lab. We may conclude from this that the unit will be heavily used and therefore this specific accessory may make it difficult to accommodate the high throughput required at the labs.They break more often than standard microtomes. A30: For Semi-Automated Microtome (Lot-25), please take into consideration the following: Line 19 refers to system efficiencies, however this line shall be adapted to read as 'Accelerated mode or equivalent solution'. Please take into consideration that all bidders are encouraged to offer fully compatible technical solutions along with alternative offers which are same or better than required. However, for components that are subject to variation (such as lines 43, 45), proposed variations shall be accepted, if requirements stipulated by the beneficiary are confirmed and differences explained and justified.
Edited on:
21-Jun-2023 15:45
Edited by:
webservice@unops.org
New clarification added: Q29: Reference made to Section III - Returnable Bidding Forms, Form C, Shipment Data: Should supplier provide information about 1 unit or total quantity of equipment in shipment dimensions (total weight, total volume, containters) columns? A29: Shipment data must be provided for the total quantity of each quoted item.
Edited on:
18-Jun-2023 12:49
Edited by:
webservice@unops.org
New clarification added: Q28: Reference made to Section III - Returnable Bidding Forms, Form D: Technical Quotation Form: Is it acceptable to submit a bid for equipmnet if it is not possible to provide a document confirming some parameters? For example, if a data sheet or a manual does not allow us to complete all lines in Form D: Technical Quotation Form. A28: If datasheets are not aligned with the technical parameters provided, manufacturer's declaration should be submitted to confirm compliance of the offered item's parameters with those requested in the Technical Quotation Form.
Edited on:
18-Jun-2023 12:46
Edited by:
webservice@unops.org
New clarification added: Q27: Please clarify whether the tender is exempt from value added tax (VAT) and customs duties, import or other taxes, or fees applicable to goods imported into Uzbekistan. A27: Reference made to RFQ Particulars, Duties and Taxes, where it is stated: "UNOPS as a UN subsidiary organ, is exempt from all direct taxes, except charges for public utility services, and is exempt from customs restrictions, duties, and charges of a similar nature in respect of articles imported or exported for its official use. All offers shall be submitted net of any direct taxes and customs duties for the imported items. Indirect taxes, such as Income Tax payable by the Contractor's staff, personnel, consultants and subcontractors that shall be providing the in-country services (delivery, assembly, installation, calibration, user training, post-sales services, etc.) shall be incorporated in their all-inclusive rates and prices."
Edited on:
18-Jun-2023 12:41
Edited by:
webservice@unops.org
New clarification added: Q26: Reference made to Form D: Technical Quotation Form, Delivery requirements –– Comparative Data Table, "Delivery place and Incoterms rules": a. FCA (Incoterms 2020) Bidder’s quoted point - should bidder quote FCA point in tender documentation or this information should be provided later by the awarded bidder? b. DPU (Incoterms 2020) Tashkent, Uzbekistan - is warehouse in Tashkent, Uzbekistan provided by the supplier or UNOPS? A26a. FCA (Incoterms 2020) point must be named by the bidder in the relevant Returnable Bidding Form. Regarding the warehouse in Tashkent, please refer to RFQ Section II: Schedule of Requirements (Rev1_of_31May23), point 9, where it is stated: Storage in Tashkent, delivery to hospitals, storage at hospitals together with associated safety and security arrangements (including adherence to the cold-chain or temperature-control regime(s) required for the storage and transportation of the consumables, where necessary) for the goods until their handover to the authorities at each hospital shall remain the responsibility of the Supplier.
Edited on:
18-Jun-2023 12:39
Edited by:
webservice@unops.org
New clarification added: Q25: Can a company with the products made in Russia participate in the bidding process? A25: Any company with the products made in any country may participate in the bidding process. However, all manufacuturer and product certification requirements must be met.
Edited on:
18-Jun-2023 12:33
Edited by:
webservice@unops.org
New clarification added: Q24: Will UNOPS provide a tax examption letter to the name of the awarded company to do the customs clearance of the equipment with the name of the end-user mentioned? A24: With the reference to RFQ requirements, Section II: Schedule of Requirements (Rev1_of_31May23), point 4: Customs clearance of the shipments shall be the responsibility of the Consignee; however, the Supplier shall have to provide the necessary documentation in a timely manner for facilitating the process.
Edited on:
18-Jun-2023 12:31
Edited by:
webservice@unops.org
New clarification added: Q23: An extention of the deadline is requested. A23: The request was accepted. The deadline for the submission of quotations is extended for 7 more days until 26 June 2023 and deadline for clarification requests was extended for 7 more days until 19 June 2023.
Edited on:
12-Jun-2023 11:41
Edited by:
webservice@unops.org
New clarification added: Q22: Reference made to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, (i)L ot 23 "Paraffin Embedding Station", Section B no. 5: - We believe that the temperature requested, up to 80ºC, is not recommended when working with paraffin as the essential properties may be damaged and consequently the efficiency of the instrument and the integrity of the tissue sample. Will you please share what is the justification for requiring 80ºC?; (ii) Lot 25 "Semi-Automated Microtome", (iia) Section B no. 19: Can you please clarify what is here required? , Section B no. 21 to 25: The selected steps in the cutting range are not adding any specific benefit to the user but redirecting the tender to a concrete brand. We would like you to reconsider this and include more standard steps in the specification. (iib) Section B no. 27 to 33: Trimming up to 700 µ. Same as above. 700 µ is not a depth of interest for microtomy but again is redirecting the tender to a concrete brand. 600 µ will enlarge the number of participants and it does not mean any benefit for the user. Also the selected steps in range are redirecting to a concrete brand; (iic) Section B no. 45: Kindly request a clarification for the addition of this accessory. We have always considered that this accessory is adequate for research, for a lab with a low throughput but never for routine sectioning in a clinical lab as the workflow is much slower. We would like to understand why it is added as a compulsory accessory. A22: The request was addressed accordingly and reflected in the updated document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV2_12June23"
Edited on:
12-Jun-2023 11:33
Edited by:
webservice@unops.org
New clarification added: Q21: Reference made to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, LOT 1 "Blood Gas & Electrolytes Analyzer", para C - please change the linearity range of the following parameters: 4.1. pH; 5.2. cNa+; 7.2. ctHb; 7.0. FMetHb. These parameters with the specified linearity range limit competition and exclude a number of manufacturers from the competition? A21: The request was addressed accordingly and reflected in the updated document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV2_12June23"
Edited on:
12-Jun-2023 11:29
Edited by:
webservice@unops.org
New clarification added: Q20: Since there are quite lots of items involved in this tender project, is it possible that the deadline of submission could be extended for a week? A20: The request was accepted. The deadline for the submission of quotations is extended for 7 more days until 26 June 2023 and deadline for clarification requests was extended for 7 more days until 19 June 2023.
Edited on:
12-Jun-2023 11:24
Edited by:
webservice@unops.org
New amendment added #4: The RFQ is further amended as follows: Deadline for the submission of quotations is extended for 7 more days until 26 June 2023;Deadline for clarification requests is extended for 7 more days until 19 June 2023;To reflect changes in Lot 1, Lot 23, Lot 25 and Lot 29, the document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV1_08June23" is hereby replaced with the updated document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV2_12June23"
Edited on:
12-Jun-2023 11:18
Edited by:
webservice@unops.org
New clarification added: Q17: As per Tender Particulars, quotation currency is mentioned as USD. According to Decree of the President of the Republic of Uzbekistan, dated September 2, 2017 No. UP-5177 «ON PRIORITY MEASURES TO LIBERALIZE MONETARY POLICY», p. 3. Establish that on the territory of the Republic of Uzbekistan: it is prohibited to make payments in foreign currency for goods (works and services), with the exception of payments through international payment cards in accordance with international practice. According to the mentioned above, can you please clarify, what currency will be used as payment currency for local bidder (supplier who is registered and making business in the Republic of Uzbekistan)? A17: As per RFQ particulars, quotations shall be submitted in USD only and the resulting contract shall be signed in USD. If the Supplier is obliged by the Law of its country of incorporation to operate only in local currency (Uzbek Sums (UZS) in the case of Uzbek companies), the USD amount of the Supplier's invoice shall be paid in the applicable local currency by applying the United Nations Operational Rates of Exchange (https://treasury.un.org/operationalrates/OperationalRates.php) applicable on the date of payment.
Edited on:
12-Jun-2023 06:37
Edited by:
webservice@unops.org
New clarification added: Q19: Will partial bids - only for some lots but not all lots be accepted? A19: As per RFP Particulars, bidders shall be allowed to quote prices for one or more lots identified in this tender. However, bidders must offer 100% of the items specified for each lot and to 100% of the quantities specified for each item of a lot. Evaluation will be done per lot.
Edited on:
10-Jun-2023 07:52
Edited by:
webservice@unops.org
New clarification added: Q18: As per Tender Particulars, quotation currency is mentioned as USD. According to Decree of the President of the Republic of Uzbekistan, dated September 2, 2017 No. UP-5177 «ON PRIORITY MEASURES TO LIBERALIZE MONETARY POLICY», p. 3. Establish that on the territory of the Republic of Uzbekistan: it is prohibited to make payments in foreign currency for goods (works and services), with the exception of payments through international payment cards in accordance with international practice. According to the mentioned above, can you please clarify, what currency will be used as payment currency for local bidder (supplier who is registered and making business in the Republic of Uzbekistan)? A18: The question was escalated yesterday to the relevant UNOPS departments. We will share the answer soon.
Edited on:
09-Jun-2023 07:11
Edited by:
webservice@unops.org
New clarification added: Q16: Are there any payments (on www.ungm.org, esourcing.unops.org or any other platform) needed from the bidder to participate in RFQ/2023/45672? A16: UNOPS does not charge any fee in connection with the expression of interest or bid submission process through www.ungm.org and/or esourcing.unops.org.
Edited on:
08-Jun-2023 13:12
Edited by:
webservice@unops.org
New clarification added: Q15: Regarding mandatory document: Balance sheets or financial statements for 2020, 2021 and 2022 (if 2022 is not available, then 2019) from document checklist. Please clarify, what information should be listed in this document? A15: The required documents and information included shall be used for assessment of the bidder's financial soundness and compliance with qualification criteria (annual turnover and liquidity rate).
Edited on:
08-Jun-2023 12:57
Edited by:
webservice@unops.org
New clarification added: Q14: Section III: Returnable Bidding Forms, Form D: Technical Quotation Form - please clarify the exact number of requested test amounts defined in the technical specification under the "H- Components, Spare Parts, Consumables" subtitle? Are these test numbers total for each lot or are they specific to each instrument? A14: All components, spare part and consumables are stipulated per item, therefore they are specific for each instrument. However, general rules and requirements also apply, as per Sections II and III of the RFQ/2023/45672
Edited on:
08-Jun-2023 12:51
Edited by:
webservice@unops.org
New clarification added: Q13: Section III: Returnable Bidding Forms, Form D: Technical Quotation Form, Lot 5, paragraph E (lines 34-36, 40) - there are requirements for workstation hardware. Will alternative offer be considered, if offered system's hardware does not meet listed requirements but will provide user with appropriate (same or better) work experience? A13: please take into consideration that all bidders are encouraged to offer fully compatible technical solutions along with alternative offers which are same or better than required. However, components that are subject to variation (such as software) are acceptable if requirements stipulated by the beneficiary are confirmed and differences justified.
Edited on:
08-Jun-2023 12:48
Edited by:
webservice@unops.org
New clarification added: Q12: Section III: Returnable Bidding Forms, Form D: Technical Quotation Form, (i) Can more specific installation requirements be provided by the end customer, e.g. gas supply for the AAS (item 9), reagents, eluents, venting systems, etc.?; (ii) please specify the requisted number of days per training; (iii) RUO or IVD devices (if IVD modules are available) should be offered?; (iv) What is the required testing procedure after installation (standard completion report, IQ/OQ, IQ/OQ based on EP or USP)? (v) What is the scope/duration of technical support expected in the years after installation? A12: (i) Partuculars are stipulated in Sections II and III of the RFQ; installation requirements will be shared between most successful bidder/contractor and the beneficiary at the time of installation. As per reagents and eluents, the bidders must comply with requirements from Section III, Form D for each item specifically, and with Section II.2d - Assembly and Calibration; (ii) As per previous response, most succesful bidders will have to arrange adequate number of days per training with the beneficiary. However, bidders are encouraged to include standard training packages into calculations/final offers; (iii) Equipment specified in RFQ/2023/45672 is aimed for clinical purposes, therefore it is IVD related; (iv) As per Section II.2d - Assembly and Calibration and Section II.2e - Online or On-Site Training Services, standard completion report based on any pharmacopeia is required; (v) Please refer to Section II.2f - Warranty and After-sales Services and Section II.2g - General Requirements.
Edited on:
08-Jun-2023 12:44
Edited by:
webservice@unops.org
New clarification added: Q11: Section III: Returnable Bidding Forms, Form D: Technical Quotation Form, lot 3 "biochemistry instrument" - we will get CE certification in the early 2024. I'd like to know whether we can participate the lot? A11: As per Section II.2b - Manufacturer & Supplier Certification, and specifically Section II.2c, please take into consideration the following: - Compliance with internationally recognized standards of quality is expected for all the offered items: Compliance and certification to Regulation (EU) 2017/745 (MDR) or Directive 93/42/EEC or FDA approval is required. However, other certificates issued in compliance with the regulatory frameworks of the founder / member countries of the Global Harmonization Task Force (GHTF) or the International Medical Device Regulators Forum (IMDRF) as a body continuing the GHTF’s mission are also acknowledged, if properly explained by the bidder (for all items offered/types of equipment).
Edited on:
08-Jun-2023 12:00
Edited by:
webservice@unops.org
New amendment added #3: The RFQ is further amended as follows:The correct Minutes of Pre-bid meeting held on 06/06/2023 for RFQ-2023-45672 is uploaded under the Documents Tab.To reflect the change in line 14 of LOT 29 "REFRIGERATED CENTRIFUGE", the document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table" is hereby replaced with the updated document titled "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table_REV1_08June23".All other requirements, terms and conditions of the RFQ remain unchanged.
Edited on:
08-Jun-2023 07:31
Edited by:
webservice@unops.org
New clarification added: Q10: Can international company participate in this RFQ? A10: Bidders from any country are eligible to participate. The only requirement is the presence in Uzbekistan through (i) representative office/branch, (ii) JV or (iii) subcontracting a locally incorporated company for handling the in-country services.
Edited on:
07-Jun-2023 10:18
Edited by:
webservice@unops.org
New clarification added: Q9: Are the devices that are not registered in Uzbekistan yet, but are about to be registered, If after winning, it is registered until the contract is concluded, acceptable?A9: We may not have fully understood the question. However, based on our understanding of the question, we would like to inform you that both registered and unregistered items can be offered. If an item is already registered, no further action for registration shall be necessary. If an item is not yet registered, the Supplier will have to initiate the registration process after signing the contract with UNOPS for the supply of the subject item. Customs clearance of the items can be done based on the initiated, but not yet completed registration processes. If registration is not completed by the time of delivery and customs clearance, registration process may be completed after distribution and installation of the subject item at all hospitals.
Edited on:
07-Jun-2023 08:38
Edited by:
webservice@unops.org
New amendment added #2: The RFQ is further amended to upload Minutes of Pre-bid meeting minutes to the Documents section.
Edited on:
07-Jun-2023 05:56
Edited by:
webservice@unops.org
New clarification added: Q8: Reference made to Section II: Schedule of Requirements - please clarify if all required items need registration in Uzpharm Control & SES. We believe the following do not need to be registered (Analytical Lab Balance, Analytical Lab Balance, CO2 Incubator, Refrigerated Centrifuge, Lab Thermostat, Rotary evaporator, Fraction Collector, Automated Polarimeter, Liquid Chromatograph, Atomic Absorption Spectrophotometer, Single-Beam Luminescent Spectrometer, Double Beam Spectrophotometer, Lab Vortex Mixer). Please check and confirm. Is there any possibility of expediting the registration procedure by the UNOPS after receiving the award? A8: Suppliers can visit the website of the State Register of Medical Products to check if the offered equipment is already registered in the Republic of Uzbekistan (link provided in the RFQ Section II - Schedule of Requirements). If equipment is not yet registered, to facilitate the customs clearance process the Supplier will have to provide evidence of application and payment for registration accompanied with the Regulatory Authority’s letter addressed to Customs. Certificate of Registration can be then issued after delivery and installation of the equipment at final destination. Application to registration can be processed online.
Edited on:
07-Jun-2023 05:37
Edited by:
webservice@unops.org
New clarification added: Q7: Section III: Returnable Bidding Forms, Form D: Technical Quotation Form, LOT 29 "REFRIGERATED CENTRIFUGE" - pelase confirm also the point 26-27 and 28 and clarify the number of tubes that should spin in the rotor? (200 pcs for 90ml and 25 ml is not a standard size)? A7: lines 26-28 refer to requirements stipulated by the beneficiary, therefore technical specifications remain the same. However, bidders can offer adequate solutions which are matching requirements, and in this case will be evaluated as compliant.
Edited on:
06-Jun-2023 11:41
Edited by:
webservice@unops.org
New clarification added: Q6: Section III: Returnable Bidding Forms, Form D: Technical Quotation Form, LOT 29 "REFRIGERATED CENTRIFUGE" - pelase confirm the specifiaction at point 14 `Operation of the rotor at ambient temperature from -5 to +600°C and relative humidity of the medium up to 80% at a temperature of +250°C`. We belive there is an imprecision in the T range. A6: the mistypo should be rephrased into: 'Operation of the rotor at ambient temperature from -5 to +60°C and relative humidity of the medium up to 80% at a temperature of +25°C`. The relevant changes will be reflected in the Revision to "RFQ-2023-45672_Section III_Returnable Bidding Forms_Form D_Technical specifications for goods – Comparative Data Table" to be uploaded later.
Edited on:
05-Jun-2023 15:14
Edited by:
webservice@unops.org
New clarification added: Q5: Reference made to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, Lot 6 "Flow Cytometer and Reagents" - (i) can laser power differ from requested mW if sufficient resolution is acheieved in sample populations? (ii) PMTs are older, less sensitive technology. Will APD detectors with higher sensitivity and improved quantum efficiency be accepted? (iii) request states a minimum of 8 channels with the ability to upgrade additional laser/decsctor lines. we would propose a 10-13 color system with additional laser at start. Would this satify the criteria? (iv) will software in languages other than Russian be considered? A5: in response to subquestions (i), (ii) and (iii) - all bidders are encouraged to offer minimal or better technical characteristics of items, therefore the request is acceptable and will be evaluated as compliant; in response to subquesiton (iv) - as per Section III, form D, software in English and/or Russian language will be evaluated as compliant.
Edited on:
05-Jun-2023 08:49
Edited by:
webservice@unops.org
New clarification added: Q4: Reference made to Section III: Returnable Bidding Forms, Form D: Technical Specifications for Goods, Lot 2 "Automated Hematology Analyzer", line 4 with the reguired number of simultaneously determined parameters at least: 18 (this is typical for 3-part Hematology analysers), but line 25 specifies requirement to 5-part white blood cell counts differentials (this is typical for 5-part Hematology analysers). For 5-part , in its turn, it is typical at least 25 parameters and high. What will be correct? Which type of Hematology analyser requested: 3-part or 5-part? A4: please take into consideration the following: Line 4 clearly states that number of parameters is at least 18; therefore all offers above this number will be considered as compliant. Line 25 clearly states that 5-part blood cell counts is requested, therefore bidders are obliged to offer 5-diff part analyser.
Edited on:
01-Jun-2023 13:13
Edited by:
webservice@unops.org
New clarification added: Q3: Is it possible to submit commercial offers in another currency (for example in Euro) and whether a local participant can submit an offer in SUM? A3: as per RFQ Particulars, the quotation currency is US dollars. Quotations received in any currency other than US Dollars shall not be considered.
Edited on:
01-Jun-2023 08:16
Edited by:
webservice@unops.org
New clarification added: Q2: Regarding the Lot 32 (Ultralow freezer), could you please let us know where the 2 units will be allocated? When checking the Table 3 "Allocation by destination", no quantity are mentionned. A2: As per Table 3 "Allocation by destination" of both "RFQ-2023-45672_Section_II_Schedule_of_Requirements" and replaced with RFQ-2023-45672_Section_II_Schedule_of_Requirements_Rev1_31May23", 2 units of Lot 2 shall be delivered to D1: Republican Specialized Scientific and practical Medical Center of Oncology and Radiology (RSSPMCOR). Address: Shifokorlar Street, Almazar District, Tashkent city.
Edited on:
31-May-2023 13:18
Edited by:
webservice@unops.org
New amendment added #1: The RFQ is amended to correct the quantities for various lots in the Table 1 of Section II: Schedule of Requirements. The document titled "RFQ-2023-45672_Section_II_Schedule_of_Requirements" is hereby replaced with the document titled "RFQ-2023-45672_Section_II_Schedule_of_Requirements_Rev1_31May23".
Edited on:
31-May-2023 08:54
Edited by:
webservice@unops.org
New clarification added: Q1: Can you please clarify the exact number of Automated tissue processors (Lot 22)? In the begging of the Schedule of requirments in Table 1 the number of units for mentioned for Lot 22 is 2, but in the same document in Table 3 for Lot 22 total number of units is 16. How many units are being procured? A1: Please refer to correct quantities for each lot as noted in Table 3 of Section II: Schedule of Requirements (PDF file) and Section III_Returnable Forms_Form C_Price Schedule Form. The typos in quantities inTable 1 of Section II: Schedule of Requirements will be addressed through Amendment of the document and uploaded later.
Edited on:
31-May-2023 04:57
Edited by:
webservice@unops.org